A versatile light microscope heating stage for biological temperatures
Authors:
- 1 June 1986
- journal article
- research article
- Published by Wiley in Journal of Microscopy
- Vol. 142 (3) , 371-374
Abstract
A light microscope heating stage suitable for biological temperatures is described. A novel feature of the design is the use of cupro-nickel foil heters (0.1 mm thick). With a temperature sensor mounted on the slide clip and a temperature controller unit, this apparatus maintains a temperature of 310K with a precision of 0.1K. The apparatus is inexpensive, reliable, not confined to use with one microscope and does not interfere significantly with the stage movement or optical alignment.Keywords
This publication has 6 references indexed in Scilit:
- Quantal secretion and response lag demonstrated in single rat neutrophilsBiochimica et Biophysica Acta (BBA) - Molecular Cell Research, 1985
- Measurement of cytosolic free Ca2+ in individual small cells using fluorescence microscopy with dual excitation wavelengthsCell Calcium, 1985
- Intracellular pH and free calcium changes in single cells using quene 1 and quin 2 probes and fluorescence microscopyFEBS Letters, 1983
- Intracellular Measurements of Ion ActivitiesAnnual Review of Biophysics and Bioengineering, 1983
- The kinetics of lysosomal degranulation of human neutrophils as measured by 9-aminoacridine quenchingBiochimica et Biophysica Acta (BBA) - Molecular Cell Research, 1983
- Optical probes of membrane potentialThe Journal of Membrane Biology, 1976