In vitro comparison of cyclic fatigue resistance of TruNatomy in single and double curvature canals compared with different nickel-titanium rotary instruments
Open Access
- 4 February 2020
- journal article
- research article
- Published by Springer Nature in BMC Oral Health
- Vol. 20 (1) , 1-8
Abstract
Background: The purpose of this study was to compare the cyclic fatigue resistance of newly developed TruNatomy instruments (TRN) in single and double (S-shaped) curvature canals with HyFlex CM (HCM), Vortex Blue (VB) and RaCe (RC) instruments.Methods: Size 20/.04 taper and size 25/0.04 of HCM, VB and RC were used. For TRN instruments, size 20/.04 taper (small) and size 26/.04 taper (prime) were used. The instruments were tested in artificial canals with double curvature (coronal curve; 60° curvature, 5 mm radius and apical curve; 70° curvature and 2 mm radius) and single curvature (60° curvature, 5 mm radius). The number of cycles to failure (NCF) was recorded. Data were statistically analyzed by Kruskal-Wallis and Dunn’s multiple comparison tests. Weibull analysis was performed on NCF data. Statistical significant was set atp < 0.05.Results: TRN and HCM revealed higher NCF compared with the other instruments for both tested sizes in single and double curvature canals (p < 0.05). TRN and HCM showed no statistically significant difference in the NCF (p > 0.05). The probability of survival was higher for HCM and TRN instruments than VB and RC instruments.Conclusions: HCM and TRN instruments were more resistant to cyclic fatigue than VB and RC instruments in single and double curvature canals. HCM and TRN instruments were anticipated to survive with higher number of cycles than the other tested instruments. RC instrument had the lowest fatigue resistance than the other instruments.Keywords
All Related Versions
This publication has 32 references indexed in Scilit:
- Evaluation of the Resistance to Cyclic Fatigue among ProTaper Next, ProTaper Universal, and Vortex Blue Rotary InstrumentsJournal of Endodontics, 2014
- Comparison of preparation of curved root canals with Hyflex CM and Revo‐S rotary nickel–titanium instrumentsInternational Endodontic Journal, 2013
- Fatigue resistance of rotary instruments manufactured using different nickel–titanium alloys: a comparative studyOdontology, 2012
- An in vitro assessment of the physical properties of novel Hyflex nickel–titanium rotary instrumentsInternational Endodontic Journal, 2012
- Alteration in the inherent metallic and surface properties of nickel–titanium root canal instruments to enhance performance, durability and safety: a focused reviewInternational Endodontic Journal, 2011
- Phase transformation behaviour and bending property of twisted nickel-titanium endodontic instrumentsInternational Endodontic Journal, 2011
- Comparison of cyclic fatigue resistance of used and new RaCe instrumentsOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2010
- Effect of heat treatment on transformation temperatures and bending properties of nickel–titanium endodontic instrumentsInternational Endodontic Journal, 2009
- Stress distribution of three NiTi rotary files under bending and torsional conditions using a mathematic analysisInternational Endodontic Journal, 2008
- Phase transformation behaviour and bending properties of hybrid nickel–titanium rotary endodontic instrumentsInternational Endodontic Journal, 2007