Nanocrystalline diamond sensor targeted for selective CRP detection: an ATR-FTIR spectroscopy study
Open Access
- 23 March 2016
- journal article
- research article
- Published by Springer Nature in Analytical and Bioanalytical Chemistry
- Vol. 408 (14) , 3675-3680
Abstract
Protein immobilization on functionalized fluorine-terminated nanocrystalline (NCD) films was studied by attenuated total reflection Fourier transform infrared (ATR-FTIR) spectroscopy using an immobilization protocol developed to specifically bind C-reactive protein (CRP). Using an ATR-FTIR spectroscopy method employing a force-controlled anvil-type configuration, three critical steps of the ex situ CRP immobilization were analyzed. First, the NCD surface was passivated by deposition of a copolymer layer consisting of polyethylene oxide and polypropylene oxide. Second, a synthetic modified polypeptide binder with high affinity to CRP was covalently attached to the polymeric film. Third, CRP dissolved in aqueous buffer in concentrations of 10–20 μg/mL was added on the functionalized NCD surface. Both the amide I and II bands, due to the polypeptide binder and CRP, were clearly observed in ATR-FTIR spectra. CRP amide I bands were extracted from difference spectra and yielded bands that agreed well with the reported amide I band of free (non-bonded) CRP in solution. Thus, our results show that CRP retains its secondary structure when it is attached to the polypeptide binders. Compared to previous IR studies of CRP in solution, about 200 times lower concentration was applied in the present study. Direct non-destructive ATR-FTIR analysis of C-reactive protein (CRP) selectively bound to functionalized nanocrystalline diamond (NCD) sensor surfaceKeywords
Funding Information
- Vetenskapsrådet (621-2014-5959)
This publication has 18 references indexed in Scilit:
- Powerful Protein Binders from Designed Polypeptides and Small Organic Molecules—A General Concept for Protein RecognitionAngewandte Chemie International Edition, 2011
- A novel interdigitated capacitor based biosensor for detection of cardiovascular risk markerBiosensors and Bioelectronics, 2009
- Quantification of C-Reactive protein in human blood plasma using near-infrared Raman spectroscopyThe Analyst, 2009
- Synthetic diamond films as a platform material for label‐free protein sensorsPhysica Status Solidi (a), 2009
- A novel ATR‐FTIR method for functionalised surface characterisationSurface and Interface Analysis, 2008
- A Novel ATR‐FTIR Approach for Characterisation and Identification of Ex Situ Immobilised SpeciesChemphyschem, 2007
- Protein-modified nanocrystalline diamond thin films for biosensor applicationsNature Materials, 2004
- Downscaling Fourier Transform Infrared Spectroscopy to the Micrometer and Nanogram Scale: Secondary Structure of Serotonin and Acetylcholine ReceptorsBiochemistry, 2003
- A study of the denaturation of human C-reactive protein in the presence of calcium ions and glycero-phosphorylcholineThermochimica Acta, 1999
- Secondary structure of the pentraxin female protein in water determined by infrared spectroscopy: effects of calcium and phosphorylcholineBiochemistry, 1992