Percutaneous Vascular Closure System Based on an Extravascular, Bioabsorbable Polyglycolic Plug (ExoSeal): Results from 1000 Patients
Authors:
Open Access
- 1 January 2014
- journal article
- Published by SAGE Publications in Clinical Medicine Insights: Cardiology
- Vol. 8s2, CMC.S15229-52
Abstract
Purpose: The objective of this prospective study was to evaluate the effectiveness and safety of a percutaneous closure system based on a polyglycolic acid plug for achieving hemostasis. Material and Methods: In this study from 2011 to 2014, a percutaneous vascular closure system (ExoSeal) was used in 1000 patients (mean age 70.6 ± 10.2 years), using antegrade and retrograde techniques within the context of an angiographic intervention. The system was used in conjunction with transfemoral approaches with a sheath size of 6F. Post the intervention (on the following day and after 6 weeks), follow-up was conducted clinically and using color-coded duplex ultrasound. Results: Immediate hemostasis was achieved in 939/1000 patients (93.9%). In the remaining 61 cases, a correct positioning of the polyglycolic acid plug was not possible because of malfunctioning of the device, massive vascular wall calcifications, postoperative scar tissue, or too steep a puncture angle. In these cases, manual compression was successful. There was one retroperitoneal bleeding requiring transfusion. Minor complications were observed (7.4% in total) with 10 pseudoaneurysm (1%), 63 inguinal hematomas (up to 3 cm; 6.3%), and 1 stenosis (0.1%). Conclusion: Safe and effective hemostasis is possible with the percutaneous ExoSeal closure system at puncture sizes of 6F.Keywords
This publication has 18 references indexed in Scilit:
- Einsatz eines neuen perkutanen Gefäßverschlusssystems nach 6F transfemoralen Zugängen: erste Ergebnisse von 100 PatientenRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2012
- Safety and Efficacy of the StarClose Vascular Closure Device in More Than 1000 Consecutive Peripheral Angioplasty ProceduresJournal of Endovascular Therapy, 2011
- Efficacy and Safety of Antegrade Common Femoral Artery Access Closure Using the Angio-Seal Device: Experience With 1889 Interventions for Critical Limb Ischemia in Diabetic PatientsJournal of Endovascular Therapy, 2010
- The StarClose Vascular Closure Device in Antegrade and Retrograde Punctures: A Single-Center ExperienceJournal of Endovascular Therapy, 2010
- A Randomized Comparison of a Novel Bioabsorbable Vascular Closure Device Versus Manual Compression in the Achievement of Hemostasis After Percutaneous Femoral ProceduresJACC: Cardiovascular Interventions, 2009
- Safety and Efficacy of the StarClose Vascular Closure System Using 7-F and 8-F Sheath Sizes:A Consecutive Single-Center AnalysisJournal of Endovascular Therapy, 2009
- Clip-based arterial haemostasis after antegrade common femoral artery punctureInternational Journal of Cardiology, 2008
- Can the Perclose suture-mediated closure system be used safely in patients undergoing diagnostic and therapeutic angiography to treat chronic lower extremity ischemia?Journal of Vascular Surgery, 2003
- A prospective study of ultrasound scan-guided thrombin injection of femoral pseudoaneurysm: A trend toward minimal medicationJournal of Vascular Surgery, 2002
- Thrombininjektion in der Behandlung des iatrogenen Aneurysma spuriumGefässchirurgie, 2002