Retrospective clinical analysis of stented vs. stentless porcine aortic bioprostheses✩
Authors:
- Mario VrandecicMVMario Vrandecic
- Fernando Antonio FantiniFFFernando Antonio Fantini
- Bayard Gontijo FilhoBFBayard Gontijo Filho
- Ozanan César de OliveiraOOOzanan César de Oliveira
- Idail Martins da Costa JúniorIJIdail Martins da Costa Júnior
- Erika VrandecicEVErika Vrandecic
- 1 July 2000
- journal article
- Published by Oxford University Press (OUP) in European Journal of Cardio-Thoracic Surgery
- Vol. 18 (1) , 46-53
Abstract
The study was designed to compare hemodynamic performance, structural failure and survival of patients undergoing aortic valve replacement (AVR) with a composite aortic stented or stentless porcine bioprosthesis. From January 1990 to June 1999, the clinical data of 725 patients undergoing AVR with stented porcine aortic bioprosthesis were reviewed. We defined two groups of patients with similar clinical characteristics: 202 patients receiving aortic stented and 205 patients stentless valves. The two patients groups were similar in age, sex, valve lesion, valve size, preoperative New York Heart Association (NYHA) class status and follow-up. The number of patients available for follow-up, excluding hospital and late mortality, reoperations and patients lost to follow-up, was 157 for the stented and 175 for the stentless group. There was a higher incidence of rheumatic heart disease in the stented (59%) vs. stentless group (44%), (P=0.003). Fewer patients had prior aortic bioprosthetic dysfunction in the stented (7.6%) compared to the stentless group (25%) (P<0.001). The mean intensive care unit stay, hospital mortality and late mortality were similar (P, NS). The total complication rate was higher in the stented (12%) than the stentless (3.4%)(P=0.005). Valve related death was higher in the stented (2.5%) than the stentless (0%) (P=0. 049). Postoperatively, the aortic effective orifice area (AEOA) was larger (P<0.001) and the transvalvular peak and mean gradients were lower in the stentless group (P<0.001). The leaflet tissue degeneration analysis was 8.0% in patients at risk for stented and 0. 6% for stentless (P=0.001). Actuarial analysis disclosed no statistical difference in patient survival between groups (P=0.18). Reoperations were less frequent in the stentless group (P=0.010). Hemodynamic benefits in the stentless group were evident and expressed by larger AEOA, lower gradients, better left ventricular remodeling with significant decrease of the left ventricular mass. Lower complication rates, lower reoperation rates, less leaflet tissue degeneration, and lower valve related mortality rates were seen in the stentless group. A controlled clinical comparison trial with longer follow-up will be required to confirm these clinical and hemodynamic benefits.Keywords
This publication has 17 references indexed in Scilit:
- Mechanisms of bioprosthetic heart valve failure: Fatigue causes collagen denaturation and glycosaminoglycan lossJournal of Biomedical Materials Research, 1999
- Twenty-year clinical experience with porcine bioprosthesesThe Annals of Thoracic Surgery, 1996
- Carpentier-edwards standard porcine bioprosthesis: Clinical performance to seventeen yearsThe Annals of Thoracic Surgery, 1995
- Porcine bioprostheses in the elderly: Clinical performance by age groups and valve positionsThe Annals of Thoracic Surgery, 1995
- Porcine aortic valve bioprostheses: A morphologic comparison of the effects of fixation pressureJournal of Biomedical Materials Research, 1990
- Microcomputer-assisted univariate survival data analysis using Kaplan-Meier life table estimatorsComputer Methods and Programs in Biomedicine, 1988
- Noninvasive estimation of valve area in patients with aortic stenosis by Doppler ultrasound and two-dimensional echocardiography.Circulation, 1985
- Problems in echocardiographic volume determinations: Echocardiographic-angiographic correlations in the presence or absence of asynergyThe American Journal of Cardiology, 1976
- HETEROLOGOUS AORTIC VALVE TRANSPLANTATIONThe Lancet, 1965
- HOMOGRAFT REPLACEMENT OF THE AORTIC VALVEThe Lancet, 1962