Proposals for the Russian quality assurance program in computed tomography
Authors:
Open Access
- 15 April 2021
- journal article
- research article
- Published by SPRI of Radiation Hygiene Prof. PV Ramzaev in Radiatsionnaya Gygiena = Radiation Hygiene
- Vol. 14 (1) , 17-33
Abstract
To ensure the quality assurance of CT-examinations, it is necessary to obtain the high-quality diagnostic information and maintain the optimal exposure levels of patients and medical staff. This paper is focused on the requirements and main aspects of quality assurance of CT-examinations, which include quality control of the equipment, methods of CT-image quality control, optimization of radiation protection, as well as management of the unintended and accidental medical exposure. The paper contains recommendations on quality control of diagnostic equipment, methods for monitoring the quality control of CT-images, values of diagnostic reference levels for the detection of abnormally high patient doses and optimization of the radiation protection of patients, as well as the recommendations for management of radiation and non-radiation accidents. All main sections of the paper represent an unified quality assurance system in computed tomography.Keywords
This publication has 20 references indexed in Scilit:
- Image quality with iterative reconstruction techniques in CT of the lungs—A phantom studyEuropean Journal of Radiology Open, 2018
- Recommendations for radiographers and radiation therapists drawn from an analysis of errors on Australian Radiation Incident RegistersJournal of Medical Radiation Sciences, 2017
- Unintended and accidental medical radiation exposures in radiology: guidelines on investigation and preventionJournal of Radiological Protection, 2017
- Practical implementation of the diagnostic reference levels concept for the common radiographic examinationsRadiatsionnaya Gygiena = Radiation Hygiene, 2017
- PATIENT DOSES IN COMPUTED TOMOGRAPHY EXAMINATIONS IN TWO REGIONS OF THE RUSSIAN FEDERATIONRadiation Protection Dosimetry, 2016
- Six iterative reconstruction algorithms in brain CT: a phantom study on image quality at different radiation dose levelsBritish Journal of Radiology, 2013
- A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise‐power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantomMedical Physics, 2013
- Image quality assessment tools for optimization of CT imagesRadiography, 2010
- The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: Practical tips, artifact examples, and pitfalls to avoidMedical Physics, 2004
- A practical method to measure the MTF of CT scannersMedical Physics, 1982