Piezocomposite transducer design and performance for high resolution ultrasound imaging transducers

Authors:

Abstract
Piezocomposite design for dedicated ultrasonic imaging applications requires precise homogenization models for predicting the electromechanical characteristics of the new material. Thus, several homogenization models have been developed. As part of this work, we applied several analytical homogenization models for piezocomposite of 2–2 and 1–3 connectivities. To validate these analytical models, a comparative study was made between various models and experimental measurements. As a result, these homogenized electromechanical properties are effectively used for the calculation and comparison of electroacoustic response for typical transducers aimed at ultrasound imaging applications. An optimal design of transducer aimed at ultrasound imaging applications is proposed as a dedicated imaging performance index, elaborated through a trade-off between sensitivity and bandwidth.

This publication has 12 references indexed in Scilit: