Declining complication rates with flow diversion of anterior circulation aneurysms after introduction of the Pipeline Flex: analysis of a single-institution series of 568 cases
- 1 December 2018
- journal article
- research article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 129 (6) , 1475-1481
Abstract
OBJECTIVE: The second-generation Pipeline embolization device (PED), Flex, has several design upgrades, including improved opening and the ability to be resheathed, in comparison with the original device (PED classic). The authors hypothesized that Flex is associated with a lower rate of major complications.METHODS: A prospective, IRB-approved, single-institution database was analyzed for all patients with anterior circulation aneurysms treated by flow diversion. The PED classic was used from August 2011 to January 2015, and the Pipeline Flex has been used since February 2015.RESULTS: A total of 568 PED procedures (252 classic and 316 Flex) were performed for anterior circulation aneurysms. The average aneurysm size was 6.8 mm. Patients undergoing treatment with the Flex device had smaller aneurysms (p = 0.006) and were more likely to have undergone previous treatments (p = 0.001). Most aneurysms originated along the internal carotid artery (89% classic and 75% Flex) but there were more anterior cerebral artery (18%) and middle cerebral artery (7%) deployments with Flex (p = 0.001). Procedural success was achieved in 96% of classic and 98% of Flex cases (p = 0.078). Major morbidity or death occurred in 3.5% of cases overall: 5.6% of classic cases, and 1.9% of Flex cases (p = 0.019). On multivariate logistic regression, predictors of major complications were in situ thrombosis (OR 4.3, p = 0.006), classic as opposed to Flex device (OR 3.7, p = 0.008), and device deployment in the anterior cerebral artery or middle cerebral artery as opposed to the internal carotid artery (OR 3.5, p = 0.034).CONCLUSIONS: Flow diversion of anterior circulation cerebral aneurysms is associated with an overall low rate of major complications. The complication rate is significantly lower since the introduction of the second-generation PED (Flex).Keywords
This publication has 23 references indexed in Scilit:
- Flow diverter device for the treatment of small middle cerebral artery aneurysmsJournal of NeuroInterventional Surgery, 2015
- International Retrospective Study of the Pipeline Embolization Device: A Multicenter Aneurysm Treatment StudyAmerican Journal of Neuroradiology, 2014
- Pipeline endovascular device for the treatment of intracranial aneurysms at the level of the circle of Willis and beyond: multicenter experienceJournal of NeuroInterventional Surgery, 2014
- Preliminary experience with the Pipeline Flex Embolization Device: technical noteJournal of NeuroInterventional Surgery, 2014
- Last-Recorded P2Y12 Reaction Units Value Is Strongly Associated with Thromboembolic and Hemorrhagic Complications Occurring Up to 6 Months after Treatment in Patients with Cerebral Aneurysms Treated with the Pipeline Embolization DeviceAmerican Journal of Neuroradiology, 2013
- The Pipeline Embolization DeviceNeurosurgery, 2013
- Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical TrialRadiology, 2013
- Immediate and follow-up results for 44 consecutive cases of small (<10 mm) internal carotid artery aneurysms treated with the pipeline embolization deviceSurgical Neurology International, 2013
- Intracranial Aneurysms: Midterm Outcome of Pipeline Embolization Device—A Prospective Study in 143 Patients with 178 AneurysmsRadiology, 2012
- Immediate procedural outcomes in 35 consecutive pipeline embolization cases: a single-center, single-user experienceJournal of NeuroInterventional Surgery, 2012