Positron Emission Mammographic Instrument: Initial Results

Authors:
  • MAMarianne Aznar
    Marianne Aznar
  • ABAlanah M. Bergman
    Alanah M. Bergman
  • CTChristopher J. Thompson
    Christopher J. Thompson
  • JRJames L. Robar
    James L. Robar
  • RLRobert Lisbona
    Robert Lisbona
  • ALAntoine Loutfi
    Antoine Loutfi
  • JGJean H. Gagnon
    Jean H. Gagnon

Abstract
Performance characteristics of a positron emission mammographic (PEM) instrument were studied. This dedicated metabolic breast imaging system has spatial resolution of 2.8-mm full width at half maximum (FWHM), coincidence resolving time of 12-nsec FWHM, and absolute efficiency of 3%. Hot spots with diameter of 16 mm in a phantom with signal-to-background activity ratio of 6:1 were distinguishable with a scanning time of 5 minutes.

This publication has 16 references indexed in Scilit: