Performance of Patients Using Different Cochlear Implant Systems: Effects of Input Dynamic Range
- 1 April 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Ear & Hearing
- Vol. 28 (2) , 260-275
Abstract
Objective: To determine, for patients who had identical levels of performance on a monosyllabic word test presented in quiet, whether device differences would affect performance when tested with other materials and in other test conditions. Design: For Experiment 1, from a test population of 76 patients, three groups (N = 13 in each group) were created. Patients in the first group used the CII Bionic Ear behind-the-ear (BTE) speech processor, patients in the second group used the Esprit3G BTE speech processor, and patients in the third group used the Tempo+ BTE speech processor. The patients in each group were matched on (i) monosyllabic word scores in quiet, (ii) age at testing, (iii) duration of deafness, and (iv) experience with their device. Performance of the three groups was compared on a battery of tests of speech understanding, voice discrimination, and melody recognition. In Experiments 2 (N = 10) and 3 (N = 10) the effects of increasing input dynamic range in the 3G and CII devices, respectively, was assessed with sentence material presented at conversational levels in quiet, conversational levels in noise, and soft levels in quiet. Results: Experiment 1 revealed that patients fit with the CII processor achieved higher scores than Esprit3G and Tempo+ patients on tests of vowel recognition. CII and Tempo+ patients achieved higher scores than Esprit3G patients on difficult sentence material presented in noise at +10 and +5 dB SNR. CII patients achieved higher scores than Esprit3G patients on difficult sentence material presented at a soft level (54 dB SPL). Experiment 2 revealed that increasing input dynamic range in the Esprit3G device had (i) no effect at conversational levels in quiet, (ii) degraded performance in noise, and (iii) improved performance at soft levels. Experiment 3 revealed that increasing input dynamic range in the CII device improved performance in all conditions. Conclusions: Differences in implant design can affect patient performance, especially in difficult listening situations. Input dynamic range and the method by which compression is implemented appear to be the major factors that account for our results.Keywords
This publication has 26 references indexed in Scilit:
- Multichannel Cochlear Implants: Relation of Histopathology to PerformanceThe Laryngoscope, 2006
- Recognition of Speech Presented at Soft to Loud Levels by Adult Cochlear Implant Recipients of Three Cochlear Implant SystemsEar & Hearing, 2004
- Effects of the Acoustical Dynamic Range on Speech Recognition with Cochlear Implants:Efectos en el rango dinámico del reconocimiento del habla con implantes coclearesInternational Journal of Audiology, 2001
- Effect of Compression Ratio on Speech Recognition and Speech-Quality Ratings With Wide Dynamic Range Compression AmplificationJournal of Speech, Language, and Hearing Research, 2000
- Effect of acoustic dynamic range on phoneme recognition in quiet and noise by cochlear implant usersThe Journal of the Acoustical Society of America, 1999
- Speech intelligibility as a function of the number of channels of stimulation for signal processors using sine-wave and noise-band outputsThe Journal of the Acoustical Society of America, 1997
- Multivariate Predictors of Audiological Success with Multichannel Cochlear ImplantsAnnals of Otology, Rhinology & Laryngology, 1993
- Stream Segregation and Peripheral ChannelingMusic Perception: An Interdisciplinary Journal, 1991
- Acoustic cues for consonant identification by patients who use the Ineraid cochlear implantThe Journal of the Acoustical Society of America, 1990
- Consonant Recognition as a Function of the Number of Channels of Stimulation by Patients Who Use the Symbion Cochlear ImplantEar & Hearing, 1989