Probe-Specific Procedure to Estimate Sensitivity and Detection Limits for 19F Magnetic Resonance Imaging
Open Access
- 11 October 2016
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 11 (10) , e0163704
Abstract
Due to low fluorine background signal in vivo, 19F is a good marker to study the fate of exogenous molecules by magnetic resonance imaging (MRI) using equilibrium nuclear spin polarization schemes. Since 19F MRI applications require high sensitivity, it can be important to assess experimental feasibility during the design stage already by estimating the minimum detectable fluorine concentration. Here we propose a simple method for the calibration of MRI hardware, providing sensitivity estimates for a given scanner and coil configuration. An experimental “calibration factor” to account for variations in coil configuration and hardware set-up is specified. Once it has been determined in a calibration experiment, the sensitivity of an experiment or, alternatively, the minimum number of required spins or the minimum marker concentration can be estimated without the need for a pilot experiment. The definition of this calibration factor is derived based on standard equations for the sensitivity in magnetic resonance, yet the method is not restricted by the limited validity of these equations, since additional instrument-dependent factors are implicitly included during calibration. The method is demonstrated using MR spectroscopy and imaging experiments with different 19F samples, both paramagnetically and susceptibility broadened, to approximate a range of realistic environments.Keywords
This publication has 41 references indexed in Scilit:
- Validating Excised Rodent Lungs for Functional Hyperpolarized Xenon-129 MRIPLOS ONE, 2013
- Imaging of Intratumoral Inflammation during Oncolytic Virotherapy of Tumors by 19F-Magnetic Resonance Imaging (MRI)PLOS ONE, 2013
- In Vivo Tracking of Human Neural Stem Cells with 19F Magnetic Resonance ImagingPLOS ONE, 2011
- Noise figure characterization of preamplifiers at NMR frequenciesJournal of Magnetic Resonance, 2011
- Fluorine (19F) MRS and MRI in biomedicineNMR in Biomedicine, 2010
- 19F MRI for quantitative in vivo cell trackingTrends in Biotechnology, 2010
- In Vivo Monitoring of Inflammation After Cardiac and Cerebral Ischemia by Fluorine Magnetic Resonance ImagingCirculation, 2008
- Multiplicative or t 1 Noise in NMR SpectroscopyApplied Magnetic Resonance, 2007
- In vivo 31P MRS of human brain at high/ultrahigh fields: a quantitative comparison of NMR detection sensitivity and spectral resolution between 4 T and 7 TMagnetic Resonance Imaging, 2006
- Quadrupolar relaxation of hyperpolarized krypton-83 as a probe for surfacesSolid State Nuclear Magnetic Resonance, 2006