Abstract
Objectives To compare the stability of a screw‐retained connection in a novel two‐piece zirconia implant to a conventional titanium‐based connection in an in vitro chewing simulation including artificial aging. Material and methods Incisor (I) and molar (M) shaped monolithic zirconia crowns were screw‐retained on either two‐piece zirconia (test) or two‐piece titanium (control) implants resulting in 4 groups of 8 samples (titanium implants with incisor shaped crowns (T‐I), titanium implants with molar shaped crowns (T‐M), zirconia implants with incisor shaped crowns (Z‐I) and zirconia implants with molar shaped crowns (Z‐M). These were subjected to artificial aging by thermal cycling (TC: 2x3000x5°C/55°C cycles of 2 min) and mechanical loading (ML: 1.2•106 cycles of 50N, f = 1 Hz). Surviving samples additionally underwent a fracture force test. Kaplan‐Meier‐plots were drawn and two‐way ANOVA was calculated taking anatomical localisation and material variables as factors. Results The mean corresponding survival times were lower for T‐M (0.86•106 ± 0.31•106 cycles) and Z‐I (0.84•106 ± 0.21•106 cycles) compared to T‐I (1.14•106 ± 0.10•106 cycles) and Z‐M (1.20•106 ± 0.10•106 cycles). In one‐way ANOVAs for survival time dependent on either location or material, no statistically significant differences could be found (location: p = .31; material: p = .62) in one‐way ANOVAs. The interaction of location and material showed significant differences (F = 21.3, p < .001). Conclusion The connection of the tested screw‐retained zirconia crowns in two‐piece zirconia implants is comparable to standard titanium implants in the specific in vitro testing.