Quantitative Assessment of Artifacts on Cardiac Magnetic Resonance Imaging of Patients With Pacemakers and Implantable Cardioverter-Defibrillators
Open Access
- 1 November 2011
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Cardiovascular Imaging
- Vol. 4 (6) , 662-670
Abstract
Background—: The safety and clinical utility of MRI at 1.5 T in patients with cardiac implantable devices such as pacemakers (PM) and implantable cardioverter-defibrillators (ICD) have been reported. This study aims to evaluate the extent of artifacts on cardiac magnetic resonance (CMR) in patients with PM and ICD (PM/ICD). Methods and Results—: A total of 71 CMR studies were performed with an established safety protocol in patients with prepectoral PM/ICD. The artifact area around the PM/ICD generator was measured in all short-axis (SA), horizontal (HLA), and vertical long-axis (VLA) SSFP cine planes. The location and extent of artifacts were also assessed in all SA (20 sectors per plane), HLA, and VLA (6 sectors per plane) late gadolinium-enhanced CMR (LGE-CMR) planes. The artifact area on cine CMR was significantly larger with ICD versus PM generators in each plane ( P <0.001, respectively). In patients with left-sided ICD or biventricular ICD systems, the percentages of sectors with any artifacts on LGE-CMR were 53.7%, 48.0%, and 49.2% in SA, HLA, and VLA planes, respectively. Patients with left-sided PM or right-sided PM/ICD had fewer artifacts. Anterior and apical regions were severely affected by artifact caused by left-sided PM/ICD generators. Conclusions—: In contrast to patients with right-sided PM/ICD and left-sided PM, the anterior and apical left ventricle can be affected by susceptibility artifacts in patients with left-sided ICD. Artifact reduction methodologies will be necessary to improve the performance of CMR in patients with left sided ICD systems.Keywords
This publication has 28 references indexed in Scilit:
- Multimodality Fusion Imaging Using Delayed-Enhanced Cardiac Magnetic Resonance Imaging, Computed Tomography, Positron Emission Tomography, and Real-Time Intracardiac Echocardiography to Guide Ventricular Tachycardia Ablation in Implantable Cardioverter-Defibrillator PatientsHeart Rhythm, 2009
- Metal-related artifacts in instrumented spine. Techniques for reducing artifacts in CT and MRI: state of the artEuropean Spine Journal, 2009
- Infarct architecture and characteristics on delayed enhanced magnetic resonance imaging and electroanatomic mapping in patients with postinfarction ventricular arrhythmiaHeart Rhythm, 2009
- Safety and Imaging Quality of MRI in Pediatric and Adult Congenital Heart Disease Patients with PacemakersPacing and Clinical Electrophysiology, 2009
- Delayed-Enhanced Magnetic Resonance Imaging in Nonischemic CardiomyopathyJACC, 2009
- How to perform magnetic resonance imaging on patients with implantable cardiac arrhythmia devicesHeart Rhythm, 2009
- Late Gadolinium Enhancement by Cardiovascular Magnetic Resonance Heralds an Adverse Prognosis in Nonischemic CardiomyopathyJACC, 2008
- Safety of Magnetic Resonance Imaging in Patients With Cardiovascular DevicesCirculation, 2007
- Clinical Utility and Safety of a Protocol for Noncardiac and Cardiac Magnetic Resonance Imaging of Patients With Permanent Pacemakers and Implantable-Cardioverter Defibrillators at 1.5 TeslaCirculation, 2006
- Strategy for Safe Performance of Extrathoracic Magnetic Resonance Imaging at 1.5 Tesla in the Presence of Cardiac Pacemakers in Non–Pacemaker-Dependent PatientsCirculation, 2006