Retrograde trans-anterior communicating artery rescue of unopened Pipeline Embolization Device with balloon dilation: complication management
Authors:
- 8 February 2014
- journal article
- case report
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 7 (2) , e7
Abstract
As the use of the Pipeline Embolization Device (PED) for the treatment of complex intracranial aneurysms rises, knowledge about complications continues to accumulate amidt a paucity of reports on techniques and rescue strategies. We describe the case of a 70-year-old woman who presented with worsening reto-orbital left-sided pain and a large cavernous aneurysm. The patient underwent endovascular treatment with PED, and there was difficulty delivering the device due to significant vascular tortuosity. This resulted in poor PED deployment as the proximal end failed to open. Increasingly aggressive strategies were attempted to open the device, which resulted in an iatrogenic carotid cavernous fistula. We were finally able to rescue the device and open its proximal end with balloon inflation after using a contralateral trans-anterior communicating artery approach and crossing the PED in a retrograde fashion. Excessive vascular tortuosity poses a genuine risk of PED malfunction and poor deployment. Although we were able to rescue the device and our patient had no permanent morbidity, difficult vascular anatomy rendered the procedure extremely complicated with dreaded complications.Keywords
This publication has 11 references indexed in Scilit:
- Current Status of Pipeline Embolization Device in the Treatment of Intracranial Aneurysms: A ReviewWorld Neurosurgery, 2013
- Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical TrialRadiology, 2013
- Safety of the Pipeline Embolization Device in Treatment of Posterior Circulation AneurysmsAmerican Journal of Neuroradiology, 2012
- Review of Reported Complications Associated with the Pipeline Embolization DeviceWorld Neurosurgery, 2012
- Pipeline Flow-Diverter Stent for Endovascular Treatment of Intracranial Aneurysms: Preliminary Experience in 20 Patients with 27 AneurysmsWorld Neurosurgery, 2011
- Retrograde Trans-Posterior Communicating Artery Snare-Assisted Rescue of Lost Access to a Foreshortened Pipeline Embolization Device: Complication ManagementOperative Neurosurgery, 2010
- Abdominal Craniectomy Implantation and Thromboembolism Prophylaxis Resulting in Wound HematomaNeurosurgery, 2010
- Treatment of Intracranial Aneurysms by Functional Reconstruction of the Parent Artery: The Budapest Experience with the Pipeline Embolization DeviceAmerican Journal of Neuroradiology, 2010
- CURATIVE ENDOVASCULAR RECONSTRUCTION OF CEREBRAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- Validity of test occlusion studies prior to internal carotid artery sacrificeHead & Neck, 1994