Pixelated CdTe detectors to overcome intrinsic limitations of crystal based positron emission mammographs
- 1 January 2013
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 8 (01) , C01030
Abstract
A positron emission mammograph (PEM) is an organ dedicated positron emission tomography (PET) scanner for breast cancer detection. State-of-the-art PEMs employing scintillating crystals as detection medium can provide metabolic images of the breast with significantly higher sensitivity and specificity with respect to standard whole body PET scanners. Over the past few years, crystal PEMs have dramatically increased their importance in the diagnosis and treatment of early stage breast cancer. Nevertheless, designs based on scintillators are characterized by an intrinsic deficiency of the depth of interaction (DOI) information from relatively thick crystals constraining the size of the smallest detectable tumor. This work shows how to overcome such intrinsic limitation by substituting scintillating crystals with pixelated CdTe detectors. The proposed novel design is developed within the Voxel Imaging PET (VIP) Pathfinder project and evaluated via Monte Carlo simulation. The volumetric spatial resolution of the VIP-PEM is expected to be up to 6 times better than standard commercial devices with a point spread function of 1 mm full width at half maximum (FWHM) in all directions. Pixelated CdTe detectors can also provide an energy resolution as low as 1.5% FWHM at 511 keV for a virtually pure signal with negligible contribution from scattered events.Keywords
This publication has 18 references indexed in Scilit:
- Energy and coincidence time resolution measurements of CdTe detectors for PETJournal of Instrumentation, 2013
- Clinical Performance of 2 Dedicated PET Scanners for Breast Imaging: Initial EvaluationJournal of Nuclear Medicine, 2012
- Imaging sensitivity of dedicated positron emission mammography in relation to tumor sizeThe Breast, 2012
- PET-Guided Breast BiopsyThe Breast Journal, 2011
- Clinical Imaging Characteristics of the Positron Emission Mammography Camera: PEM Flex Solo IIJournal of Nuclear Medicine, 2009
- Standardized Uptake Values of Normal Breast Tissue with 2-Deoxy-2-[F-18]Fluoro-d-glucose Positron Emission Tomography: Variations with Age, Breast Density, and Menopausal StatusMolecular Imaging and Biology, 2006
- Instrumentation optimization for positron emission mammographyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- Results of preliminary clinical trials of the positron emission mammography system PEM-I: a dedicated breast imaging system producing glucose metabolic images using FDG.2000
- Positron emission mammography (PEM): a promising technique for detecting breast cancerIEEE Transactions on Nuclear Science, 1995
- Feasibility study for positron emission mammographyMedical Physics, 1994