Use of the HemoCue ® near patient testing device to measure the concentration of haemoglobin in suction fluid at elective Caesarean section*
Authors:
Open Access
- 11 April 2008
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 63 (5) , 531-534
Abstract
Summary: We have investigated the suitability of the HemoCue ® photometer to measure the concentration of haemoglobin in suction fluid obtained at elective caesarean section in 30 women. Laboratory analysis was used as a gold standard against which values generated by the HemoCue were compared. We used the method of Bland and Altman to analyse the data. The bias and the limits of agreement were −0.013 and −0.39 to 0.36 mg.dl −1 respectively, indicating a good level of agreement. Mean (SD) total blood loss calculated using these data, combined with the weight of the swabs, was consistently greater than clinical estimation: 768 (496) ml versus 506 (249) ml respectively (p < 0.001). We have found that the HemoCue near patient testing device may be used to estimate blood loss accurately in the suction fluid obtained at elective Caesarean section.Keywords
This publication has 5 references indexed in Scilit:
- Use of a new HemoCue system for measuring haemoglobin at low concentrationsClinical and Laboratory Haematology, 2001
- Blood Loss During Transurethral Resection of the Prostate as Measured by the Hemocue PhotometerScandinavian Journal of Urology and Nephrology, 1993
- Intra‐operative blood loss during elective lower segment caesarean sectionBJOG: An International Journal of Obstetrics & Gynaecology, 1992
- Evaluation of "HemoCue," a new device for determining hemoglobin.Clinical Chemistry, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986