High-Field and Low-Field MR Imaging of Superior Glenoid Labral Tears and Associated Tendon Injuries

Authors:

Abstract
OBJECTIVE. The purpose of this study was to determine the performance characteristics of high-field and low-field MR imaging for the diagnosis of a glenoid superior labral anteroposterior (SLAP) tear.MATERIALS AND METHODS. High-field (n = 46) or low-field (n = 21) MR imaging was performed on 41 patients with SLAP tears and 26 patients with normal superior labra. The superior labrum was classified into one of four types on the basis of patterns of intralabral signal intensity. The relative frequency of rotator cuff tears and long head of the biceps tendinopathy was also assessed.RESULTS. For the diagnosis of SLAP tear, the sensitivity of high-field MR imaging was 90% (95% confidence interval = 74%, 98%), specificity was 63% (35%, 85%), and accuracy was 80% (66%, 91%). The sensitivity of low-field MR imaging was 64% (31%, 89%), specificity was 70% (35%, 93%), and accuracy was 67% (43%, 85%). A branched linear or stellate focus of abnormal intralabral signal intensity was associated with a SLAP tear in 86% o...