Abstract
In this work an ultra-low frequency electric field probe capable of measuring time varying gradient coil induced electric fields in a tissue mimicking material was constructed, calibrated and validated. This probe consisted of a 2 cm short dipole antenna followed by an instrumentation. The designed amplifier had a gain of 100 and a cut-off frequency of 31 kHz. The probe was fabricated on a 4-layer printed circuit board (PCB), FR4 of thickness 1.57 mm and a copper thickness of 35 μm. A waterproofed 3D probe holder was used to cover the electronics and expose the dipole tip to the saline solution. To calibrate this probe in a known environment a parallel-plate conductor was used. The calibrated probe was validated by measuring the electric field in a saline filled cylindrical phantom placed within a cylindrically symmetric test coil. In order to do the validation, simulations were performed using an in-house tool developed in MATLAB and a validated tool in Sim4Life.
Funding Information
  • Natural Sciences and Engineering Research Council of Canada
  • Ontario Research Fund
  • CMC Microsystems

This publication has 12 references indexed in Scilit: