The extravascular implantable cardioverter-defibrillator: characterization of anatomical parameters impacting substernal implantation and defibrillation efficacy
Authors:
Open Access
- 18 October 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 24 (5) , 762-773
Abstract
The aim of this study is to provide a thorough, quantified assessment of the substernal space as the site of extravascular implantable cardioverter-defibrillator (ICD) lead placement using computed tomography (CT) scans and summarizing adverse events and defibrillation efficacy across anatomical findings. Subcutaneous ICDs are an alternative to transvenous defibrillators but have limitations related to ICD lead distance from the heart. An alternative extravascular system with substernal lead placement has the potential to provide defibrillation at lower energy and pacing therapies from a single device. A multi-centre, non-randomized, retrospective analysis of 45 patient CT scans quantitatively and qualitatively assessing bony, cardiac, vascular, and other organ structures from two human clinical studies with substernal lead placement. Univariate logistic regression was used to evaluate 15 anatomical parameters for impact on defibrillation outcome and adjusted for multiple comparisons. Adverse events were summarized. Substernal implantation was attempted or completed in 45 patients. Defibrillation testing was successful in 37 of 41 subjects (90%) using ≥10 J safety margin. There were two intra-procedural adverse events in one patient, including reaction to anaesthesia and an episode of transient atrial fibrillation during ventricular fibrillation induction. Anatomical factors associated with defibrillation failure included large rib cage width, myocardium extending very posteriorly, and a low heart position in the chest (P-values <0.05), though not significant adjusting for multiple comparisons. Retrospective analysis demonstrates the ability to implant within the substernal space with low intra-procedural adverse events and high defibrillation efficacy despite a wide range of anatomical variability.Keywords
Funding Information
- Medtronic PLC
- Boston Scientific Corp.
This publication has 20 references indexed in Scilit:
- Longevity of the Subcutaneous Implantable DefibrillatorCirculation: Arrhythmia and Electrophysiology, 2015
- 2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac deathEP Europace, 2015
- The Safety and Efficacy of Mediastinoscopy When Performed by General Thoracic SurgeonsThe Annals of Thoracic Surgery, 2014
- World Medical Association Declaration of HelsinkiJAMA, 2013
- An Entirely Subcutaneous Implantable Cardioverter–DefibrillatorNew England Journal of Medicine, 2010
- A comparison of outcomes after robotic open extended thymectomy for myasthenia gravis☆European Journal of Cardio-Thoracic Surgery, 2007
- The superior epigastric artery does not pass through Larrey's space ( trigonum sternocostale )Surgical and Radiologic Anatomy, 2003
- Clinical Predictors of Defibrillation Thresholds with an Active Pectoral Pulse Generator Lead SystemPacing and Clinical Electrophysiology, 2002
- Manubrium-sparing median sternotomy as a uniform approach for cardiac operations.2000
- CT anatomy of the internal mammary vessels: importance in planning percutaneous transthoracic procedures.American Journal of Roentgenology, 1990