Comparison of four 31P single‐voxel MRS sequences in the human brain at 9.4 T
Open Access
- 11 January 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 85 (6) , 3010-3026
Abstract
Purpose: In this study, different single‐voxel localization sequences were implemented and systematically compared for the first time for phosphorous MRS (31P‐MRS) in the human brain at 9.4 T.Methods: Two multishot sequences, image‐selected in vivo spectroscopy (ISIS) and a conventional slice‐selective excitation combined with localization by adiabatic selective refocusing (semiLASER) variant of the spin‐echo full intensity–acquired localized spectroscopy (SPECIAL‐semiLASER), and two single‐shot sequences, semiLASER and stimulated echo acquisition mode (STEAM), were implemented and optimized for 31P‐MRS in the human brain at 9.4 T. Pulses and coil setup were optimized, localization accuracy was tested in phantom experiments, and absolute SNR of the sequences was compared in vivo. The SNR per unit time (SNR/t) was derived and compared for all four sequences and verified experimentally for ISIS in two different voxel sizes (3 × 3 × 3 cm3, 5 × 5 × 5 cm3, 10‐minute measurement time). Metabolite signals obtained with ISIS were quantified. The possible spectral quality in vivo acquired in clinically feasible time (3:30 minutes, 3 × 3 × 3 cm3) was explored for two different coil setups.Results: All evaluated sequences performed with good localization accuracy in phantom experiments and provided well‐resolved spectra in vivo. However, ISIS has the lowest chemical shift displacement error, the best localization accuracy, the highest SNR/t for most metabolites, provides metabolite concentrations comparable to literature values, and is the only one of the sequences that allows for the detection of the whole 31P spectrum, including β–adenosine triphosphate, with the used setup. The SNR/t of STEAM is comparable to the SNR/t of ISIS. The semiLASER and SPECIAL‐semiLASER sequences provide good results for metabolites with long T2.Conclusion: At 9.4 T, high‐quality single‐voxel localized 31P‐MRS can be performed in the human brain with different localization methods, each with inherent characteristics suitable for different research issues.Keywords
Funding Information
- Cancer Prevention and Research Institute of Texas (RR180056)
This publication has 67 references indexed in Scilit:
- Quantitative imaging of energy expenditure in human brainNeuroImage, 2012
- SPECIAL semi‐LASER with lipid artifact compensation for 1H MRS at 7 TMagnetic Resonance in Medicine, 2012
- Localized 1H NMR spectroscopy in different regions of human brain in vivo at 7 T: T2 relaxation times and concentrations of cerebral metabolitesNMR in Biomedicine, 2011
- Semi-LASER localized dynamic31P magnetic resonance spectroscopy in exercising muscle at ultra-high magnetic fieldMagnetic Resonance in Medicine, 2011
- Neurochemical alterations in spinocerebellar ataxia type 1 and their correlations with clinical statusMovement Disorders, 2010
- Spectroscopic imaging with improved gradient modulated constant adiabaticity pulses on high-field clinical scannersJournal of Magnetic Resonance, 2010
- Optimal phased-array combination for spectroscopyMagnetic Resonance Imaging, 2008
- In vivo 31P MRS of human brain at high/ultrahigh fields: a quantitative comparison of NMR detection sensitivity and spectral resolution between 4 T and 7 TMagnetic Resonance Imaging, 2006
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988
- Localized proton spectroscopy using stimulated echoesJournal of Magnetic Resonance (1969), 1987