Abstract
Flow Cytometry (FCM) is still considered to be the method of choice for accurate CD4 enumeration. However, the use of FCM in developing countries is problematic due to their cost and complexity. Lower-cost technologies have been introduced. We evaluated CyFlow Counter together with its lyophilized reagents, and Pima CD4 in high-temperature area, using FACSCount as reference. Whole blood samples were consecutively collected by venipuncture from 111 HIV+ patients and 17 HIV-negative donors. CD4 T-cell enumeration was performed on CyFlow Counter, Pima CD4 and FACSCount. CyFlow Counter and Pima CD4 systems showed good correlation with FACSCount (slope of 0.82 and 0.90, and concordance ρc of 0.94 and 0.98, respectively). CyFlow Counter showed absolute or relative biases (LOA) of −63 cells/mm3 (−245 to 120) or −9.8% (−38.1 to 18.4) respectively, and Pima CD4 showed biases (LOA) of −30 cells/mm3 (−160 to 101) or −3.5% (−41.0 to 33.9%). CyFlow Counter and Pima CD4 showed respectively 106.7% and 105.9% of similarity with FACSCount. According to WHO-2010 ART initiation threshold of 350 cells/mm3, CyFlow Counter and Pima CD4 showed respectively sensibility of 100% and 97%, and specificity of 91% and 93%. CyFlow Counter and Pima CD4 were strongly correlated (slope of 1.09 and ρc of 0.95). These alternative systems showed good agreement with bias of 33 cells/mm3 (−132 to 203) or 6.3% (−31.2 to 43.8), and similarity of 104.3%. CyFlow Counter using CD4 easy count kit-dry and Pima CD4 systems can accurately provide CD4 T-cell counts with acceptable agreement to those of FACSCount.Keywords
This publication has 39 references indexed in Scilit:
- Evaluation of PIMA™® Point of Care Technology for CD4 T Cell Enumeration in KenyaPLOS ONE, 2013
- Evaluation of Portable Point-of-Care CD4 Counter with High Sensitivity for Detecting Patients Eligible for Antiretroviral TherapyPLOS ONE, 2012
- Performance evaluation of the Pima™ point‐of‐care CD4 analyser using capillary blood sampling in field tests in South AfricaJournal of the International AIDS Society, 2012
- Utility of the point of care CD4 analyzer, PIMA, to enumerate CD4 counts in the field settings in IndiaAIDS Research and Therapy, 2012
- Evaluation of the Partec Flow Cytometer against the BD FACSCalibur System for Monitoring Immune Responses of Human Immunodeficiency Virus-Infected Patients in ZimbabweClinical and Vaccine Immunology, 2007
- The Partec CyFlow Counter® could provide an option for CD4+ T-cell monitoring in the context of scaling-up antiretroviral treatment at the district level in MalawiTransactions of the Royal Society of Tropical Medicine and Hygiene, 2006
- A Microchip CD4 Counting Method for HIV Monitoring in Resource-Poor SettingsPLOS Medicine, 2005
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part IClinical Chemistry and Laboratory Medicine (CCLM), 1983
- A Coefficient of Agreement for Nominal ScalesEducational and Psychological Measurement, 1960