Abstract
To investigate the biomechanical properties of the percutaneous compression plating system for intertrochanteric hip fractures. A biomechanical laboratory investigation on human cadaveric upper femora was conducted. Biomechanical laboratory. Sixteen femora from cadavers of patients aged 60 to 85 years. An intertrochanteric fracture was performed along the intertrochanteric line and fixed with the percutaneous compression plating system. Postfixation and postcycling bending and torsional stiffnesses, and load to failure at 0, 7, and 25° of adduction. The normalized postfixation bending and torsional stiffnesses in the neutral position were 65% and 59%, respectively, and higher with adduction. After percutaneous compression plating fixation, bending stiffness increased to 72% following cyclic loading (p < 0.05). It was comparable to that of similar fixation constructs in the literature, whereas the torsional stiffness was somewhat higher. The mode of failure was distal femoral neck sagging, indicating good femoral head fixation. No varus displacement and cutouts were observed. Our biomechanical data indicate that the percutaneous compression plating provides adequate bending stiffness and torsional stability. With sliding capability this fixation leads to controlled fracture impaction, which is further enhanced by cyclic loading.Keywords
This publication has 31 references indexed in Scilit:
- Treatment of Unstable Peritrochanteric Fractures in Elderly Patients with a Compression Hip Screw or with the Vandeputte (VDP) Endoprosthesis: A Prospective Randomized StudyJournal of Orthopaedic Trauma, 1995
- Functional outcome after intertrochanteric fractures of the femur: does the implant matter? A prospective study of 100 consecutive casesInjury, 1993
- Experimentelle Untersuchungen zur Schwächung der Stabilität des Trochanter major durch laterale FensterungEuropean Journal of Trauma and Emergency Surgery, 1993
- The Incidence of Fracture of the Proximal Femur in Two Million Canadians From 1972 to 1984Clinical Orthopaedics and Related Research, 1991
- Trochanteric FracturesClinical Orthopaedics and Related Research, 1990
- The Fixation and Prognosis of Trochanteric FracturesClinical Orthopaedics and Related Research, 1990
- Torsion and bending analysis of internal fixation techniques for femoral neck fractures: The role of implant design and bone densityJournal of Orthopaedic Research, 1987
- THE INCREASING INCIDENCE OF FRACTURES OF THE PROXIMAL FEMUR: AN ORTHOPAEDIC EPIDEMICThe Lancet, 1983
- Incidence of hip FracturesActa Orthopaedica, 1980
- A Clinical and Biomechanical StudyClinical Orthopaedics and Related Research, 1980