Effect of different γ‐irradiation doses on cytotoxicity and material properties of porous polyether‐urethane polymer
Authors:
- 18 July 2006
- journal article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 80B (2) , 415-423
Abstract
Biomaterials respond to sterilization methods differently. Steam sterilization might decrease the performance of thermoplastic polyether‐urethane (TPU); however, the effect of different γ‐radiation doses on this polymer is contradictory in present literature. The purpose of this study was to investigate the differences between irradiative doses in comparison with steam sterilization on a porous TPU scaffold produced by a new processing method. No significant differences in the surface chemical structure were found with attenuated total reflectance Fourier transform infrared spectroscopy (ATR‐FTIR) analysis when comparing with the sterilization methods. The molecular weight (Mw) had a net increase from 11.5 ± 0.039 to 13.2 ± 0.072 kDa by γ‐sterilization from 10 to 60 kGy. The samples that were irradiated (>60 kGy) had also an increase in polydispersity index (PDI; 1.45 ± 0.007) in comparison with the nonsterile ones (1.31 ± 0.017), which indicate branching. Liquid chromatography/mass spectroscopy (LC/MS) analysis showed that there was a correlation between the concentration of the breakdown product, methyl dianiline, and cytotoxicity. The concentration of this compound was found to be four times higher in steam‐sterilized sample (1.3 ± 0.01 ppb) compared with that of the polymer sample γ‐sterilized at 10 kGy (0.3 ± 0.01 ppb). The cytotoxicity of TPU was found to decrease with higher radiation doses, and was significantly higher for the steam‐sterilized samples. It is recommended that TPU produced with the described foaming method should be sterilized by γ‐irradiation at 25 kGy or higher doses. © 2006 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2007Keywords
This publication has 41 references indexed in Scilit:
- Effect of processing route and acetone pre-treatment on the biostability of pellethane materials used in medical device applicationsBiomaterials, 2005
- Residual formaldehyde on plastic materials and medical equipment following low-temperature steam and formaldehyde sterilizationJournal of Hospital Infection, 2005
- NMR analysis of γ-radiation induced degradation of halthane-88 polyurethane elastomersPolymer Degradation and Stability, 2003
- Radiation crosslinking of LDPE and HDPE with 5 and 10 MeV electron beamsEuropean Polymer Journal, 2001
- Thermoanalytical investigations of polyurethanes for medical purposesThermochimica Acta, 1998
- Tissue response to commercial silicone and polyurethane elastomers after different sterilization proceduresBiomaterials, 1996
- Sterilisation of implantable devicesClinical Materials, 1994
- Variations between biomer lots. 2: The effect of differences between lots on in vitro enzymatic and oxidative degradation of a commercial polyurethaneJournal of Biomedical Materials Research, 1993
- Radiation sterilization of polymeric implant materialsJournal of Biomedical Materials Research, 1988
- Determination of extractable methylene dianiline in thermoplastic polyurethanes by HPLCJournal of Biomedical Materials Research, 1984