Development of HEATHER for Cochlear Implant Stimulation Using a New Modeling Workflow
- 22 October 2014
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 62 (2) , 728-735
Abstract
The current conduction pathways resulting from monopolar stimulation of the cochlear implant were studied by developing a human electroanatomical total head reconstruction (namely, HEATHER). HEATHER was created from serially sectioned images of the female Visible Human Project dataset to encompass a total of 12 different tissues, and included computer-aided design geometries of the cochlear implant. Since existing methods were unable to generate the required complexity for HEATHER, a new modeling workflow was proposed. The results of the finite-element analysis agree with the literature, showing that the injected current exits the cochlea via the modiolus (14%), the basal end of the cochlea (22%), and through the cochlear walls (64%). It was also found that, once leaving the cochlea, the current travels to the implant body via the cranial cavity or scalp. The modeling workflow proved to be robust and flexible, allowing for meshes to be generated with substantial user control. Furthermore, the workflow could easily be employed to create realistic anatomical models of the human head for different bioelectric applications, such as deep brain stimulation, electroencephalography, and other biophysical phenomena.Keywords
Funding Information
- Australian Research Council (LP0776938)
- Australian Postgraduate Award (APA)
This publication has 37 references indexed in Scilit:
- Subject Specific Modelling of Electrical Conduction in the Body: A Case StudyJournal of Biomimetics, Biomaterials and Tissue Engineering, 2011
- Effects of local tissue conductivity on spherical and realistic head modelsAustralasian Physics & Engineering Sciences in Medicine, 2010
- Predicted current densities in the brain during transcranial electrical stimulationClinical Neurophysiology, 2006
- A realistic three dimensional FEM of the human headPhysiological Measurement, 2001
- Realistic computer modelling of electric and magnetic fields of human head and torsoParallel Computing, 1998
- Effect of conductivity uncertainties and modeling errors on EEG source localization using a 2-D modelIEEE Transactions on Biomedical Engineering, 1998
- The electrical conductivity of human cerebrospinal fluid at body temperatureIEEE Transactions on Biomedical Engineering, 1997
- Influence of tissue resistivities on neuromagnetic fields and electric potentials studied with a finite element model of the headIEEE Transactions on Biomedical Engineering, 1997
- Thickness and resistivity variations over the upper surface of the human skullBrain Topography, 1993
- The specific resistance of biological material—A compendium of data for the biomedical engineer and physiologistMedical & Biological Engineering & Computing, 1967