Abstract
Using a two‐Gaussian model for the radial distribution of the emitted Auger electrons, it is shown that the resolving power (point‐to‐point resolution) is governed by the incident probe diameter d , while the minimum detectable dimension d m may be less than d . The convolution/deconvolution problem of Auger profiles is also considered with the transfer function of scanning Auger microscopy (SAM). Auger profiles across various flat distributions (single stripe, i.e. single quantum well; isolated rectangular details; stripe arrays; i.e. superlattices; and linear concentrations at non‐abrupt interfaces) are estimated showing the distortion of the experimental profiles by the radial distribution of Auger electrons. For small details, the apparent concentration looks very different from the true one and a two‐point analysis procedure is suggested for improving the quantification of SAM at high lateral resolution as an alternative to deconvolution of the Auger profiles. The same analysis may be applied to other analytical microscopies particularly to electron probe microanalysis of a specimen having a surface compsition different from that of the bulk.