Early Outcomes With Direct Flow Medical Versus First‐Generation Transcatheter Aortic Valve Devices: A Single‐Center Propensity‐Matched Analysis

Authors:

Abstract
Objectives: To compare early outcomes between Direct Flow Medical (DFM) and 1st‐generation transcatheter aortic valve replacement (TAVR) devices.Background: The effectiveness of new‐generation TAVR devices compared with earlier generations is unclear.Methods and Results: Overall, 496 patients treated between November 2007 and March 2014 were included: 44 (8.7%) treated with DFM, 179 (36.9%) with Medtronic CoreValve (MCV) and 273 (54.4%) with Edwards Sapien / Sapien XT (SXT) valve. Because of differences in baseline characteristics, propensity‐score matching among patients treated in the same time span (March 2012–March 2014) was performed. The primary objective was the VARC‐2 device success composite endpoint. Propensity‐score matching identified 41 patients in each group. A higher device success composite endpoint was observed in DFM (DFM 97.6% vs. MCV 65.9% vs. SXT 92.7%; P < 0.001). This was attributed to a significantly lower incidence of moderate‐to‐severe post‐procedural aortic regurgitation (AR; 2.4% vs. 22% vs. 7.3%; P < 0.001), lower rate of valve embolization (0% vs. 7.3% vs. 0%; 0.041) and need for a 2nd valve implantation (0% vs. 7.3% vs. 0%; P = 0.041).Conclusions: DFM was associated with higher rates of device success and a lower incidence of post‐procedural AR. New generation devices appear to address several of the limitations of earlier generation devices. (J Interven Cardiol 2015;28:583–593)