In vitro detection of simulated apical root perforation with two electronic apex locators.
Authors:
Open Access
- 20 February 2010
- journal article
- Vol. 5 (1) , 23-6
Abstract
The aim of this study was to evaluate the accuracy of two electronic apex locators (Smarpex and NovApex) in detecting apical perforation. After access cavity preparations, the working length was determined by the first examiner before and after perforation with a stereomicroscope by introducing a K-file size 10 into the canal and size 25 K-file up to the perforation site, respectively. The specimens were embedded in a 17-well plastic box containing alginate. The root canals were irrigated with chlorhexidine 0.2% (CHX) through a 27-gauge needle. Two examiners measured the root canal length twice and the mean value was calculated. The data were analyzed using Kolmogorov-Smirnov and ANOVA tests. The results obtained with each Electronic Apex Locator were compared with the corresponding control length. The statistical analysis showed reliable accuracies in detecting the perforation site for the two experimental electronic apex locators (60% and 80% for NovApex and Smarpex, respectively). Although no significant difference was shown between Smarpex and NovApex, these results suggest that electronic apex locators can effectively and reproducibly detect root canal perforations.Keywords
This publication has 23 references indexed in Scilit:
- Early flaring and crown-down shaping influences the first file bind to the canal apical thirdOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2008
- Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root canals of extracted upper molarsOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2008
- Immunohistochemical expression of MMPs 1, 7, and 26 in syndrome and nonsyndrome odontogenic keratocystsOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2008
- Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teethInternational Endodontic Journal, 2008
- Accuracy of two different apex locators in primary teeth with and without root resorptionClinical Oral Investigations, 2007
- Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2007
- A comparison between two electronic apex locators: an ex vivo investigationInternational Endodontic Journal, 2007
- Influence of cervical preflaring on apical file size determinationInternational Endodontic Journal, 2005
- Reliability of different electronic apex locators to detect root perforations in vitroInternational Endodontic Journal, 1997
- Determination of location of root perforations by electronic apex locatorsOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 1996