Mitigating Stress Shielding in Dorr C Femurs via Additive Manufacturing: A Proof-of-Concept Numerical Analysis
Authors:
- Roberta CromiRCRoberta Cromi
- Francesca BertiFBFrancesca Berti
- Matteo GavazzoniMGMatteo Gavazzoni
- Luigi La BarberaLBLuigi La Barbera
- Dalila Di PalmaDPDalila Di Palma
- Sara MaggioniSMSara Maggioni
- Jacopo MeniniJMJacopo Menini
- Massimo FranceschiniMFMassimo Franceschini
- Stefano FolettiStefano Foletti
- Tomaso VillaTVTomaso Villa
Open Access
Abstract
Bone resorption secondary to stress shielding is a leading cause of hip implant failure, primarily due to the stiffness mismatch between the femur and the prosthesis. Although anatomical stem designs generally provide improved load transfer, Dorr type C femurs often require straight stems to ensure adequate primary stability. This work presents a systematic approach to designing a straight, additively manufactured porous titanium hip stem aimed at minimizing stress shielding. The lattice architecture is customized to replicate the mechanical properties of bone based on patient-specific femoral CT scans. The performance of the resulting porous implant is numerically assessed under simplified physiological gait loading conditions. The implant behavior is evaluated through a homogenization strategy to model the lattice structure, significantly reducing the computational effort and making the methodology easily replicable. Compared to its full counterpart, the porous design achieves a significant reduction in predicted bone loss, suggesting that the proposed framework is a promising proof of concept for patient-specific implants. While further experimental validation and larger cohort studies are required, these findings highlight the potential of mechanically tunable porous structures to mitigate the stress shielding phenomenon in anatomical conditions such as Dorr type C femurs, which require straight stems.Keywords
Funding Information
- National Plan for NRRP Complementary Investments (PNC0000003)
This publication has 58 references indexed in Scilit:
- Lattice Structures for Bone Replacement: The Intersection of Bone Biomechanics, Lattice Design, and Additive ManufacturingAdvanced Materials Technologies, 2026
- Review and assessment of various lattice structure homogenizationMultiscale and Multidisciplinary Modeling, Experiments and Design, 2025
- The cementless taper wedge vs. fit-and-fill stem in primary total hip arthroplasty: risk of stem-related complication differs across Dorr typesArchives of orthopaedic and trauma surgery, 2024
- Stress shielding effect after total hip arthroplasty varies between combinations of stem design and stiffness—a comparing biomechanical finite element analysisInternational Orthopaedics, 2023
- A comprehensive analysis of bio-inspired design of femoral stem on primary and secondary stabilities using mechanoregulatory algorithmBiomechanics and Modeling in Mechanobiology, 2020
- Occurrence, risk factors, and outcomes of bone cement implantation syndrome after hemi and total hip arthroplasty in cancer patientsJournal of Surgical Oncology, 2019
- Fully porous 3D printed titanium femoral stem to reduce stress-shielding following total hip arthroplastyJournal of Orthopaedic Research, 2016
- What Are the Causes for Failures of Primary Hip Arthroplasties in France?Clinical Orthopaedics and Related Research, 2013
- Effects of material properties of femoral hip components on bone remodelingJournal of Orthopaedic Research, 1992
- Adaptive bone-remodeling theory applied to prosthetic-design analysisJournal of Biomechanics, 1987