Cell Injection Microrobot Development and Evaluation in Microfluidic Chip
Authors:
- 1 December 2018
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-7
Abstract
We propose an innovative design of microrobot, which can achieve donor cell suction, delivery and injection in a mammalian oocyte on microfluidic chip. The microrobot body contains a hollow space that produces suction and ejection forces for injection of cell nuclei using a nozzle at the tip of the robot. Specifically, a controller changes the hollow volume by balancing the magnetic and elastic forces of the membrane, and along with motion of stages in the XY plane. A glass capillary attached at the tip of the robot contains the nozzle is able to absorb and inject cell nuclei. The microrobot provides three degrees of freedom and generates micronewton forces. We demonstrate the effectiveness of the proposed microrobot through an experiment of absorption and ejection of 20 $\mu$ m particles from the nozzle using magnetic control in a microfluidic chip.
Keywords
This publication has 20 references indexed in Scilit:
- Microrobot with passive diamagnetic levitation for microparticle manipulationsJournal of Applied Physics, 2017
- On-chip microfluid induced by oscillation of microrobot for noncontact cell transportationApplied Physics Letters, 2017
- Accurate dispensing system for single oocytes using air ejectionBiomicrofluidics, 2013
- Driving method of microtool by horizontally arranged permanent magnets for single cell manipulationApplied Physics Letters, 2010
- The synchronization of superparamagnetic beads driven by a micro-magnetic ratchetLab on a Chip, 2010
- Powerful actuation of magnetized microtools by focused magnetic field for particle sorting in a chipBiomedical Microdevices, 2010
- Magnetostatic interactions and forces between cylindrical permanent magnetsJournal of Magnetism and Magnetic Materials, 2009
- Manipulation of biological samples using micro and nano techniquesIntegrative Biology, 2009
- On the dynamics of magnetically driven elastic filamentsJournal of Fluid Mechanics, 2006
- Magnetically-driven biomimetic micro pumping using vorticesLab on a Chip, 2004