Abstract
Objectives: To analyze the incidence of malfunction of the da Vinci robotic system in a single center and to provide potential solutions.Methods: A total of 400 patients underwent da Vinci robotic urological surgery at Taichung Veterans General Hospital in Taichung, Taiwan, from December 2005 to April 2011. Episodes of malfunction of the robotic system were analyzed by period of operation, type of procedure, type of malfunction and management of the event.Results: Overall, 14 cases of malfunction occurred (3.5% of the entire series). Among them, five (1.25%) occurred before the surgery and nine (2.25%) intraoperatively. Operative procedures included radical prostatectomy, bilateral pelvic lymph node dissection, dismembered pyeloplasty, partial nephrectomy, nephroureterectomy, and radical and partial cystectomies. Areas of malfunctions included the robotic arm system and joint (11/14), optical system (1/14), power system and connector (1/14), endoscopic instrument (1/14), and software (1/14). In 10 cases, the failure was recoverable, whereas in four cases there was a critical failure, requiring a conversion to standard laparoscopy in three of them, and the rescheduling of the surgery in one case.Conclusions: The da Vinci robotic system is extremely reliable for use in urology. Malfunction is rare and the risk of critical failure is very low. Managing mechanical failure before or during the surgery is the key to maintaining the safety of patients undergoing robotic surgical procedures.Keywords
This publication has 11 references indexed in Scilit:
- The learning curve for reducing complications of robotic‐assisted laparoscopic radical prostatectomy by a single surgeonBJU International, 2010
- Robotic surgery in urologic oncology: gathering the evidenceExpert Review of Pharmacoeconomics & Outcomes Research, 2010
- Robotic‐assisted laparoscopic radical prostatectomy: Learning curve of first 100 casesInternational Journal of Urology, 2010
- Incidencias y complicaciones en nuestras primeras 250 prostatéctomias radicales robóticasActas Urológicas Españolas, 2010
- Malfunction of the da Vinci Robotic System During Robot-Assisted Laparoscopic Prostatectomy: An International SurveyJournal of Endourology, 2010
- A Unique Instrumental Malfunction during Robotic ProstatectomyYonsei Medical Journal, 2010
- Complications and incidences in our first 250 robotic radical prostatectomiesActas Urológicas Españolas (English Edition), 2010
- Failure and Malfunction of da Vinci Surgical Systems During Various Robotic Surgeries: Experience From Six Departments at a Single InstituteUrology, 2009
- Robotic Equipment Malfunction During Robotic Prostatectomy: A Multi-institutional StudyJournal of Endourology, 2008
- Da Vinci Robot Error and Failure Rates: Single Institution Experience on a Single Three-Arm Robot Unit of More than 700 Consecutive Robot-Assisted Laparoscopic Radical ProstatectomiesJournal of Endourology, 2007