Determination of chemical oxygen demand in fresh waters using flow injection with on-line UV-photocatalytic oxidation and spectrophotometric detection
- 22 December 2004
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 130 (2) , 227-232
Abstract
A flow injection manifold incorporating UV-photocatalytic oxidation for the determination of chemical oxygen demand (COD) in freshwater is reported. The method utilises the UV-photocatalytic oxidation of organic compounds instead of conventional heating (used in the standard method), with acidified potassium permanganate as the oxidant. Sodium oxalate, D-glucose and potassium hydrogen phthalate were used as COD standards. A 100 µL sample solution was injected into a 0.3 mol L−1 H2SO4 carrier stream containing 0.1 mol L−1 (NH4)2SO4, merged with a permanganate solution (8 × 10−4 mol L−1) and passed through a 250 cm FEP (fluoroethylene polymer) photo-reaction coil wound around a 15 W UV lamp. The sample throughput was 30 h−1, with an LOD (blank plus 3σ) of 0.5 mg COD L−1 and a linear range of 0.5–50 mg COD L−1 ( D-glucose, r2 = 0.9966). The method had good precision with relative standard deviations of 2.7% at 5 mg COD L−1 and 1.2% at 20 mg COD L−1 (n = 12) for glucose. Results for a COD certified reference material (QC Demand Quality Control Standard) were in good agreement with the certified COD value. Recovery from Tamar River water samples for all three COD standards was 83.0–111.0% and the COD values determined were in good agreement with those of a permanganate index reference method.Keywords
This publication has 26 references indexed in Scilit:
- Study on photocatalytic oxidation for determination of chemical oxygen demand using a nano-TiO2–K2Cr2O7 systemAnalytica Chimica Acta, 2004
- Photocatalytic sensor for the determination of chemical oxygen demand using flow injection analysisAnalytica Chimica Acta, 2001
- Chemical oxygen demand determination in environmental waters by mixed-acid digestion and single sweep polarographyAnalytica Chimica Acta, 2000
- Alternative Method for Determination of the Chemical Oxygen Demand for Colloidal Polymeric WastewaterBulletin of Environmental Contamination and Toxicology, 1997
- The coulometric determination of chemical oxygen demandElectroanalysis, 1997
- UV-photooxidation as pretreatment step in inorganic analysis of environmental samplesAnalytica Chimica Acta, 1996
- Determination of chemical oxygen demand in aqueous environmental samples by segmented flow-injection analysisAnalytica Chimica Acta, 1992
- Chemical oxygen demand determination in well and river waters by flow-injection analysis using a microwave oven during the oxidation stepAnalytica Chimica Acta, 1992
- Determination of photochemically produced hydroxyl radicals in seawater and freshwaterMarine Chemistry, 1990
- Use of flow injection analysis based on iodometry for automation of dissolved oxygen (Winkler method) and chemical oxygen demand (dichromate method) determinationsAnalytical and Bioanalytical Chemistry, 1988