Non-Enzymatic Glucose Biosensor Based on Highly Pure TiO2 Nanoparticles
Open Access
- 11 May 2021
- journal article
- research article
- Published by MDPI AG in Biosensors
- Vol. 11 (5) , 149
Abstract
This study proposes a non-enzymatic glucose sensor fabricated by synthesizing high-purity TiO2 nanoparticles in thermal plasma and depositing it directly on a substrate and then depositing chitosan–polypyrrole (CS-PPy) conductive polymer films by electrochemical method. The structural properties of the deposited TiO2 nanoparticles were analyzed by X-ray diffraction (XRD) and dynamic light scattering (DLS) system. The chemical composition and structural properties of the TiO2 nanoparticle layer and the conductive polymer films were confirmed by X-ray photoelectron spectroscopy (XPS) spectra and scanning electron microscope (SEM). The glucose detection characteristics of the fabricated biosensor were determined by cyclic voltammetry (CV). CS-PPy/TiO2 biosensor showed high sensitivity of 302.0 µA mM−1 cm−2 (R2 = 0.9957) and low detection limit of 6.7 μM. The easily manufactured CS-PPy/TiO2 biosensor showed excellent selectivity and reactivity.Keywords
Funding Information
- National Research Foundation of Korea (2018R1D1A1B07040898, 2018R1D1A1B07050522)
This publication has 36 references indexed in Scilit:
- Nanomaterials for electrochemical non-enzymatic glucose biosensorsRSC Advances, 2012
- Ultrasonic electrodeposition of platinum nanoflowers and their application in nonenzymatic glucose sensorsElectrochimica Acta, 2012
- Adsorption of Glucose Oxidase onto Single-Walled Carbon Nanotubes and Its Application in Layer-By-Layer BiosensorsAnalytical Chemistry, 2009
- New Nanostructured TiO2 for Direct Electrochemistry and Glucose Sensor ApplicationsAdvanced Functional Materials, 2008
- Electrochemical Glucose BiosensorsChemical Reviews, 2007
- Bioelectrocatalytical glucose oxidation with phenoxazine modified glucose oxidaseElectrochemistry Communications, 2004
- Carbon paste electrodes modified with enzymes, tissues, and cellsElectroanalysis, 1995
- Electrical connection of enzyme redox centers to electrodesThe Journal of Physical Chemistry, 1992
- The Enzyme ElectrodeNature, 1967
- ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERYAnnals of the New York Academy of Sciences, 1962