Selective suppression of artifact‐generating echoes in cine DENSE using through‐plane dephasing
- 11 October 2006
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 56 (5) , 1126-1131
Abstract
In displacement‐encoded imaging with stimulated echoes (DENSE), tissue displacement is encoded in the phase of the stimulated echo. However, three echoes generally contribute to the acquired signal (the stimulated echo, the complex conjugate of the stimulated echo, and an echo due to T1 relaxation). It is usually desirable to suppress all except the stimulated echo, since otherwise the additional echoes will cause displacement measurement errors. Ideally, suppression of the artifact‐generating echoes would be independent of time, T1, and displacement‐encoding frequency, and would not require additional acquisitions. In this study through‐plane gradients were used to selectively dephase artifact‐generating echoes without causing significant signal loss of the stimulated echo. A cine DENSE sequence was modified to include dephasing gradients and perform complementary spatial modulation of magnetization (CSPAMM). For single‐acquisition cine DENSE using dephasing alone, artifact suppression was similar to CSPAMM with two acquisitions. The use of dephasing with CSPAMM required two acquisitions, but demonstrated greater artifact suppression than CSPAMM alone or dephasing alone. Magn Reson Med, 2006.Keywords
This publication has 14 references indexed in Scilit:
- Displacement‐encoded cardiac MRI using cosine and sine modulation to eliminate (CANSEL) artifact‐generating echoesMagnetic Resonance in Medicine, 2004
- Complementary displacement‐encoded MRI for contrast‐enhanced infarct detection and quantification of myocardial function in miceMagnetic Resonance in Medicine, 2004
- Myocardial Tissue Tracking with Two-dimensional Cine Displacement-encoded MR Imaging: Development and Initial EvaluationRadiology, 2004
- In vivo study of microcirculation in canine myocardium using the IVIM method†Magnetic Resonance in Medicine, 2003
- High-Resolution Strain Analysis of the Human Heart with Fast-DENSEJournal of Magnetic Resonance, 1999
- DENSE: Displacement Encoding with Stimulated Echoes in Cardiac Functional MRIJournal of Magnetic Resonance, 1999
- Stimulated Anti-Echo Selection in Spatially Localized NMR SpectroscopyJournal of Magnetic Resonance, 1999
- Limitations of stimulated echo acquisition mode (steam) techniques in cardiac applicationsMagnetic Resonance in Medicine, 1995
- Diffusion and field-gradient effects in NMR Fourier spectroscopyThe Journal of Chemical Physics, 1974
- Nuclear Induction in Inhomogeneous FieldsPhysical Review B, 1955