Development of a high‐stability water vapor radiometer
Author:
- Alan B. TannerATAlan B. Tanner
- 1 March 1998
- journal article
- Published by American Geophysical Union (AGU) in Radio Science
- Vol. 33 (2) , 449-462
Abstract
Preliminary design details and laboratory test results of a prototype microwave radiometer operating near 20 GHz are presented. Radiometric stability to 10 millikelvin on time scales from 100 to 10,000 s is the goal for the final system. The use of noise diodes as receiver gain calibrators is examined, and test results showing 30 parts per million stability per day in the intercomparison of noise diodes are presented. Temperature control of the radiometer is also examined, and the design of a solid state temperature control system capable of 1 mK stability near room temperature is presented. A new design for a mechanically actuated Dicke switch is also presented. The results of these laboratory developments are applied to a prototype radiometer. Early indications are that design goals are being met in the prototype. Future confirmation of the system's theoretical performance will only be possible with the completion of the final radiometer units.Keywords
This publication has 6 references indexed in Scilit:
- New model‐based Bayesian inversion algorithm for the retrieval of wet troposphere path delay from radiometric measurementsRadio Science, 1998
- A test of water vapor radiometer-based troposphere calibration using very long baseline interferometry observations on a 21-km baselineRadio Science, 1996
- A New Instrument for the Determination of Radio Path Delay due to Atmospheric Water VaporIEEE Transactions on Geoscience and Remote Sensing, 1985
- A Statistical Measure for the Stability of Solid State Noise SourcesIEEE Transactions on Microwave Theory and Techniques, 1977
- An S-Band Radiometer Design with High Absolute PrecisionIEEE Transactions on Microwave Theory and Techniques, 1974
- Statistics of atomic frequency standardsProceedings of the IEEE, 1966