Electromagnetic Interference from Wireless Video-Capsule Endoscopy on Implantable Cardioverter-Defibrillators
Authors:
- 1 April 2007
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 30 (4) , 472-475
Abstract
The purpose of this study was to evaluate potential interference between the Pill-cam video-capsule and implanted cardio-defibrillators (ICD) in both in vitro and in vivo test environments. Phase I consisted of in vitro testing utilized 6 ICD models. Each was placed in a saline gel bath in conjunction with a tool designed to emit the same wave length as the Pill-Cam (the Test Cap). Tests were performed at both the manufacturer's nominal and most sensitive settings with the test probe placed at 1, 5, 10, and 15 cm from three different points. There were emissions of 10-, 30-, and 60- second duration at each location. Phase II was the in vivo study utilizing patients with implanted ICDs for standard clinical reasons who were undergoing evaluation with the Test Cap. In Phase I, 864 tests were performed involving the 6 ICDs. There was totally normal behavior in 5 of the devices. The Biotronik Belos (Berlin, Germany) demonstrated oversensing and delivery of inappropriate therapy. In phase II, 6 patients underwent a total number of 288 tests (48 in each patient). There were 4 men and 2 women, mean age 61 (33-77) and none demonstrated any adverse interactions. In patients with devices identical to those evaluated that showed no interference, the use of the Pill-Cam is safe. However, in patients with a Biotronik Belos ICD (or any other nontested ICD) the use of the Pill-Cam should be done only in-hospital, after suspended ICD therapy and in conjunction with close monitoring.Keywords
This publication has 12 references indexed in Scilit:
- Effects of Capsule Endoscopy on Cardiac PacemakersEndoscopy, 2005
- Safety of Wireless Capsule Endoscopy in Patients with Implantable Cardiac DefibrillatorsAmerican Journal of Gastroenterology, 2005
- LETTER TO THE EDITORPacing and Clinical Electrophysiology, 2005
- Electromagnetic interference with implantable cardiac pacemakers by video capsuleGastrointestinal Endoscopy, 2005
- Clinical Usefulness of the Endoscopic Video Capsule as the Initial Intestinal Investigation in Patients with Obscure Digestive Bleeding: Validation of a Diagnostic Strategy Based on the Patient Outcome after 12 MonthsEndoscopy, 2004
- Safety of capsule endoscopy in patients with pacemakersGastrointestinal Endoscopy, 2004
- Interference with Cardiac Pacemakers by Cellular TelephonesNew England Journal of Medicine, 1997
- Electromagnetic Interference of Pacemakers by Mobile PhonesPacing and Clinical Electrophysiology, 1996
- Evaluation of Outpatient Arrhythmias and Pacemakers Utilizing Transtelephonic Monitoring DevicesAnnals of Noninvasive Electrocardiology, 1996
- Intermittent pacemaker dysfunction caused by digital mobile telephonesJACC, 1996