Prospective multicenter international study on the outcomes of a newly developed self‐approximating lumen‐apposing metallic stent for drainage of pancreatic fluid collections and endoscopic necrosectomy
- 25 July 2019
- journal article
- research article
- Published by Wiley in Digestive Endoscopy
- Vol. 32 (3) , 391-398
Abstract
Background A novel self‐approximating lumen apposing metallic stent (LAMS) (Niti‐S Spaxus, Taewoong Medical, Korea) has recently become available. The aim of this study was to evaluate the outcomes for drainage of pancreatic fluid collections (PFC). Methods This was a prospective international multi‐centered study conducted in 6 high volume institutions across Asia. Consecutive patients suffering from pancreatic pseudocyst or walled‐off pancreatic necrosis (WOPN) requiring EUS‐guided drainage were recruited. Outcomes included technical and clinical success, adverse events, procedural events, interventions through the stent and recurrence rates. Results Between August 2016 and November 2017, 59 patients were recruited to the study. 39 patients (66.1%) had WOPN and the mean (S.D.) size of the PFC's were 11.5 (5.1) cm. The technical and clinical success rates were 100%. The mean (S.D.) procedural time was 35.0 (17.2) minutes. The 16mm stents were used in 66.1% of the patients. 54 sessions of necrosectomy were performed with the sent in situ in 17 patients. Stent‐related adverse events (AE) was 6.8%. 3 patients (5%) suffered from bleeding after stenting and 1 required angiographic embolization. Two patients (3.3%) suffered from recurrence during a mean (S.D.) follow‐up time of 325.6 (355.5) days. There were no differences in outcomes between those with pseudocysts or WOPN except for the duration of hospital stay (P = 0.012). Conclusion The use of a self‐approximating LAMS for drainage of PFC's was safe and effective. Endoscopic necrosectomy could be performed through the stent with ease. The device was associated with a low rate of stent‐related AE's.Keywords
This publication has 27 references indexed in Scilit:
- Endoscopic ultrasonography‐guided transmural drainage of walled‐off pancreatic necrosis: Comparison between a specially designed fully covered bi‐flanged metal stent and multiple plastic stentsDigestive Endoscopy, 2016
- Lumen‐apposing metal stents for drainage of pancreatic fluid collections: When and for whom?Digestive Endoscopy, 2016
- EUS-guided choledochoduodenostomy for malignant distal biliary obstruction using a lumen-apposing fully covered metal stent after failed ERCPSurgical Endoscopy, 2016
- Systematic review comparing endoscopic, percutaneous and surgical pancreatic pseudocyst drainageWorld Journal of Gastrointestinal Endoscopy, 2016
- A rare adverse event resulting from the use of a lumen-apposing metal stent for drainage of a pancreatic fluid collection: “the buried stent”Gastrointestinal Endoscopy, 2015
- Prospective evaluation of endoscopic ultrasonography-guided double-balloon-occluded gastrojejunostomy bypass (EPASS) for malignant gastric outlet obstructionGut, 2015
- A newly designed fully covered metal stent for lumen apposition in EUS-guided drainage and access: a feasibility study (with videos)Gastrointestinal Endoscopy, 2014
- Classification of acute pancreatitis—2012: revision of the Atlanta classification and definitions by international consensusGut, 2012
- Ex vivo comparative study using the Endolifter® as a traction device for enhancing submucosal visualization during endoscopic submucosal dissectionSurgical Endoscopy, 2012
- A lexicon for endoscopic adverse events: report of an ASGE workshopGastrointestinal Endoscopy, 2010