Patient safety concept for multichannel transmit coils
Authors:
Open Access
- 29 October 2007
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 26 (5) , 1315-1321
Abstract
Purpose: To propose and illustrate a safety concept for multichannel transmit coils in MRI based on finite‐differences time‐domain (FDTD) simulations and validated by measurements.Materials and Methods: FDTD simulations of specific absorption rate (SAR) distributions in a cylindrical agarose phantom were carried out for various radio frequency (RF) driving conditions of a four‐element coil array. Additionally, maps of transmit amplitude, signal phase, and temperature rise following RF heating were measured by MRI.Results: Quantitative agreement was achieved between simulated and measured field distributions, thus validating the numerical modeling. When applying the same RF power to each element of the coil array but systematically varying the RF phase between its elements, the maximum of the SAR distribution was found to vary by a factor of about 15.Conclusion: Our results demonstrate that current RF safety approaches are inadequate to deal with the new challenge of multichannel transmit coils. We propose a new concept based on a systematic investigation of the parameter space for RF phases and amplitudes. In this way the driving conditions generating the highest local SAR values per unit power can be identified and appropriately considered in the RF safety concept of a given MRI system. J. Magn. Reson. Imaging 2007;26:1315–1321.Keywords
This publication has 21 references indexed in Scilit:
- Parallel RF transmission with eight channels at 3 TeslaMagnetic Resonance in Medicine, 2006
- Experimental analysis of parallel excitation using dedicated coil setups and simultaneous RF transmission on multiple channelsMagnetic Resonance in Medicine, 2005
- Magnetic resonance temperature imagingInternational Journal of Hyperthermia, 2005
- Electromagnetic perspective on the operation of RF coils at 1.5–11.7 TeslaMagnetic Resonance in Medicine, 2005
- Radiofrequency power deposition utilizing thermal imagingMagnetic Resonance in Medicine, 2004
- Temperature and SAR calculations for a human head within volume and surface coils at 64 and 300 MHzJournal of Magnetic Resonance Imaging, 2004
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Analysis of wave behavior in lossy dielectric samples at high fieldMagnetic Resonance in Medicine, 2002
- Performance and use of current sheet antennae for RF-hyperthermia of a phantom monitored by 3 tesla MR-thermographyInternational Journal of Hyperthermia, 2002