Scanning Protocol Choice Affects Optical Coherence Tomography (OCT-3) Measurements
Authors:
- Rabia Gurses-OzdenRGRabia Gurses-Ozden
- Manchima MakornwattanaMMManchima Makornwattana
- Roberto VessaniRVRoberto Vessani
- Jeffrey M. LiebmannJLJeffrey M. Liebmann
- Celso TelloCTCelso Tello
- Robert RitchRRRobert Ritch
- 1 April 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 13 (2) , 142-144
Abstract
To determine the optimal OCT-3 scanning protocol. Normal subjects, ocular hypertensive, and open-angle glaucoma patients were enrolled. The “fast” RNFL scan protocol consists of 3 consecutive circular scans, each containing 256 A-scans obtained in a single 1.92 second session. The “regular” RNFL scan protocol contains 3 separate circular scans, each of which consists of 512 A-scans and is obtained in 1.28 seconds (total 3.84 seconds). A customized 256-point scan was created from the 512-point “regular” RNFL scan by extracting the RNFL measurements from every second point. Mean RNFL thickness measurements and mean coefficients of variation (CV) were calculated for each of the scan types. Thirty-one eyes with visual field defects (31 glaucoma eyes) and 30 eyes with normal fields (10 normal and 20 ocular hypertensive eyes) were enrolled. There was no difference in RNFL thickness or CV between fast, regular, or customized RNFL scans in the eyes with VF defects (P > 0.05 for all comparisons). Although eyes with normal fields showed no difference in CV between fast, regular, and customized RNFL scans (P > 0.05 for all comparisons), there was a thicker RNFL measurement using the fast RNFL scan (P = 0.01). Optimal sampling density for OCT-3 assessment of parapapillary RNFL thickness is not more than 256 axial determinations. Although both scanning protocols have good reproducibility, the scanning protocol should remain constant during serial OCT-3 evaluation.Keywords
This publication has 7 references indexed in Scilit:
- The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layerBritish Journal of Ophthalmology, 2001
- Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography11The authors have no financial interest in the Optical Coherence Tomography technology.Ophthalmology, 2000
- Increasing Sampling Density Improves Reproducibility of Optical Coherence Tomography MeasurementsJournal of Glaucoma, 1999
- Retinal nerve fiber layer thickness in human eyesGraefe's Archive for Clinical and Experimental Ophthalmology, 1999
- Retinal Nerve Fiber Layer Thickness in Normal Human EyesOphthalmology, 1996
- Reproducibility of Nerve Fiber Layer Thickness Measurements Using Optical Coherence TomographyOphthalmology, 1996
- Optical Coherence TomographyScience, 1991