Analysis of fumonisin B 1 by negative‐ion thermospray mass spectrometry

Authors:

Abstract
Thermospray mass spectrometry was used to analyze fumonisin B 1 (FB 1 ) from two corn samples that were known to contain fumonisins, and from a FB 1 ‐spiked meat sample. The mass spectrometer was operated in the negative‐ion discharge‐on, filament‐off, mode with the fragmentor set at 200 V. Selected‐ion monitoring (SIM) was used for quantitative analysis. The mobile phase was CH 3 CN (containing 1% CICH 2 CN, v/v)+0.1M ammonium acetate, pH 3.65 (65:35, v/v). Quantitative analysis was done by integrating the peak areas traced by ion currents of 5 selected ions in SIM, and by the method of standard additions. Responses to the selected ions were linear within the range of 2‐200 ng. Fumonisins were easily identified in the corn and muscle tissue matrices without any apparent matrix interferences and were detectable down to 0.1 μ/mL (2 ng injected).