Observations on Shear-Induced Hemolysis
Authors:
- Richard D. OffemanRORichard D. Offeman
- Michael C. WilliamsMWMichael C. Williams
- 1 January 1979
- journal article
- research article
- Published by Taylor & Francis in Biomaterials, Medical Devices, and Artificial Organs
- Vol. 7 (3) , 393-420
Abstract
New data were reported on chemical and mechanical factors affecting the hemolysis of [human] blood in a rotating-parallel-disk apparatus used in biomaterials studies described elsewhere. The hemolytic potential for blood correlated with plasma components LDH [lactate dehydrogenase] and total triglycerides existing prior to shear. Antibiotics suppressed hemolysis, unrelated to bacterial action. The kinetics of measured Hb release were analyzed in terms of a red cell population having a distribution of fragilities, sample storage age and possible Hb adsorption. Blood/gas interfacial effects were examined by varying the gas and the extent of interface, and were minimal under normal testing conditions. Several geometric and dynamic parameters were varied to illustrate the complexities of engineering design and scale-up.Keywords
This publication has 4 references indexed in Scilit:
- Red Blood Cell Damage by Shear StressBiophysical Journal, 1972
- A device to study the degradation of blood in a shear field with controlled wall effectsJournal of Biomedical Materials Research, 1971
- Hemolysis in simple shear flowsAIChE Journal, 1970
- On the flow between a rotating and a stationary diskJournal of Fluid Mechanics, 1968