A Monte Carlo study of IMRT beamlets in inhomogeneous media
Authors:
- Andrew O. JonesAJAndrew O. Jones
- Indra J. DasIDIndra J. Das
- Frederick L. JonesFJFrederick L. Jones
- 5 February 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (3) , 296-300
Abstract
Intensity Modulated Radiation Therapy (IMRT) has trended toward smaller multiple radiation fields thereby increasing the resolution of the intensity map. Vendors have introduced multileaf collimators with beam apertures of 0.5 cm and less. The beam characteristics of these smaller fields have not been adequately assessed, especially in the presence of inhomogeneities. Most dosimetric devices have significant limitations due to finite size, dose rate, and energy dependence. We studied the effect of inhomogeneities on small beamlets. The 6, 15, and 24 MV beams were modeled using the EGSnrc Monte Carlo code. Point source beams of circular field sizes 0.5, 1.0, 3.0, 5.0, and 10 cm were simulated in a water phantom at 100 SSD. A 3 cm inhomogeneity of lung tissue was incorporated between 3 and 6 cm in the phantom. The depth dose curves and profiles were compared by beam size and density of the inhomogeneity. The Monte Carlo simulations show that for small fields a marked dose decrease in the presence of low-density media due to the lack of lateral electronic equilibrium is observed. As the density and field size increase, the dose reduction is less pronounced and for the 10 cm field there is an increased dose as expected due to lack of attenuation. This data suggests that current TPS may dramatically over- or underestimate the dose in inhomogeneous media for small field sizes that are used for IMRT.Keywords
This publication has 14 references indexed in Scilit:
- Dose distribution of narrow beam irradiation for small lung tumorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapyPhysics in Medicine & Biology, 2000
- The impact of electron transport on the accuracy of computed doseMedical Physics, 2000
- Comparative dosimetry in narrow high-energy photon beamsPhysics in Medicine & Biology, 2000
- Dose perturbation caused by high-density inhomogeneities in small beams in stereotactic radiosurgeryPhysics in Medicine & Biology, 1998
- Influence of air inhomogeneities in radiosurgical beamsMedical Dosimetry, 1997
- Dosimetric characterization of radiosurgical beams with a diamond detectorMedical Physics, 1995
- Implications of tissue heterogeneity for radiosurgery in head and neck tumorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995
- Doses on the central axes of narrow 6‐MV x‐ray beamsMedical Physics, 1990
- Measurements of dose distributions in small beams of 6 MV X-raysPhysics in Medicine & Biology, 1987