Incidence and predictors of pacemaker reprogramming: potential consequences for remote follow-up

Authors:

Abstract
Remote follow-up (FU) enables to cope with the expanding number of pacemaker (PM) FU. Although remote FU offers comparable monitoring options to in-office FU, reprogramming of device settings is not available, thereby imposing a potentially important restriction to the applicability of remote FU. The aim of this study was to assess in a large cohort of bradycardia PM recipients, the incidence of PM reprogramming during long-term FU and its predictors, to judge the possibilities for remote FU. Between 2003 and 2010 all in-office FU of 1517 bradycardia PM recipients included in the FOLLOWPACE study were recorded. Only 24.5% of all 13 258 recorded FU visits >3 months after implantation were visits-with-reprogramming (VWRs), occurring in 1158 patients (79%). Fifty percent of patients were free of reprogramming at 9 months, and 29% at 24 months. Using multivariable binary logistic regression analysis, the following patient characteristics were predictive for frequent PM reprogramming, defined as >3 VWRs during 3 year FU: age, a history of atrial arrhythmias, PM complication 80%) do not need frequent reprogramming, suggesting a potential benefit of using remote FU to reduce the number of unnecessary in-office visits.