A precise, reproducible method for measuring ultrasound probe slice thickness using a Gammex 403 phantom
- 14 February 2019
- journal article
- research article
- Published by SAGE Publications in Ultrasound
- Vol. 27 (3) , 148-155
Abstract
A precise, reproducible method for measuring ultrasound probe slice thickness has been developed for linear array ultrasound probes. The method used a custom built jig to draw the probe along the surface of a Gammex 403 phantom, with the image plane parallel to the filaments within the phantom. Still images at 0.5 mm intervals are saved for post-processing using in-house software. Slice thickness measurements with a precision of 0.1 mm are obtained. The method was shown to give reproducible estimates of probe slice thickness at several depths to within 0.4 mm during repeat tests. The method was able to provide information about the slice thickness of different sections of the probe face. It is expected that the method can quantify changes in probe performance due to lens wear or replacement over time that may elude both in-plane and in-air reverberation-based tests. A total of 18 linear probes were tested across eight centres, including six specialist vascular ultrasound centres.Keywords
This publication has 8 references indexed in Scilit:
- Recognising small image quality differences for ultrasound probes and the potential of misdiagnosis due to undetected side lobesUltrasound, 2017
- A multicentre survey of the condition of ultrasound probesUltrasound, 2016
- The Resolution Integral – a tool for characterising the performance of diagnostic ultrasound scannersUltrasound, 2014
- BMUS guidelines for the regular quality assurance testing of ultrasound scanners by sonographersUltrasound, 2013
- The Automated Assessment of Ultrasound Scanner Lateral and Slice Thickness Resolution: Use of the Step ResponseUltrasound in Medicine & Biology, 2009
- Real‐time B‐mode ultrasound quality control test procedures. Report of AAPM Ultrasound Task Group No. 1Medical Physics, 1998
- Estimation of ultrasound beam width in the elevation (section thickness) plane.Radiology, 1991
- Slice thickness measurements.Journal of Ultrasound in Medicine, 1988