Characterization of ultrasound elevation beamwidth artifacts for prostate brachytherapy needle insertion
- 20 December 2011
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 39 (1) , 246-256
Abstract
Purpose: Ultrasound elevation beamwidth leads to image artifacts and uncertainties in localizing objects (such as a surgical needle) in ultrasoundimages. The authors examined the clinical significance of errors caused by elevation beamwidth artifacts and imaging parameters in needle insertion procedures. Methods: Beveled prostate brachytherapy needles were inserted through all holes of a grid template under real-time transrectal ultrasound(TRUS) guidance. The needle tip position as indicated by the TRUSimage was compared to their observed physical location. A new device was developed to measure the ultrasound elevation beamwidth. Results: Imaging parameters of the TRUS scanner have direct impact on the localization error ranging from 0.5 up to 4 mm. The smallest localization error was observed laterally close to the center of the grid template and axially within the beam’s focal zone. Largest localization error occurs laterally around both sides of the grid template and axially within the beam’s far field. The authors also found that the localization errors vary with both lateral and elevation offsets. Conclusions: The authors found properly adjusting the TRUSimaging settings to lower the ultrasound gain and power effectively minimized the appearance of elevation beamwidth artifacts and in turn reduced the localization errors of the needle tip.Keywords
This publication has 21 references indexed in Scilit:
- A compact mechatronic system for 3D ultrasound guided prostate interventionsMedical Physics, 2011
- Imaging of implant needles for real-time HDR-brachytherapy prostate treatment using biplane ultrasound transducersMedical Physics, 2009
- US ArtifactsRadioGraphics, 2009
- Brachytherapy needle deflection evaluation and correctionMedical Physics, 2005
- prostate brachytherapy: a review of current practiceJournal of Radiotherapy in Practice, 2004
- Test Object for Measurement of Section Thickness at USRadiology, 1999
- SIDE LOBES AND GRATING LOBES ARTIFACTS IN ULTRASOUND IMAGINGVeterinary Radiology & Ultrasound, 1997
- Estimation of ultrasound beam width in the elevation (section thickness) plane.Radiology, 1991
- Sonographic artifacts and their origins.American Journal of Roentgenology, 1991
- The importance of ultrasonic side-lobe artifacts.Radiology, 1982