Improved Synchrony and Respiratory Unloading by Neurally Adjusted Ventilatory Assist (NAVA) in Lung-Injured Rabbits
Authors:
- 1 March 2007
- journal article
- Published by Springer Nature in Pediatric Research
- Vol. 61 (3) , 289-294
Abstract
With increasing pressure support ventilation (PSV), a form of pneumatically triggered ventilation, there can be an increase in wasted inspiratory efforts (neural inspiratory efforts that fail to trigger the ventilator). With neurally adjusted ventilatory assist (NAVA), a mode of ventilation controlled by the electrical activity of the diaphragm (EAdi), synchrony should be maintained at high levels of assist. The aim of this study was to evaluate the response to increasing levels of PSV and NAVA on synchrony and diaphragm unloading in lung-injured rabbits. Animals were ventilated on PSV or NAVA in random order, each at three levels. We measured neural and ventilator respiratory rates, EAdi, transdiaphragmatic pressure (Pdi), and tidal volume (Vt). At low PSV, 95% of neural efforts were triggered, compared with high PSV, where only 66% of the neural efforts were triggered. During NAVA, all neural efforts were triggered, regardless of level. Increasing NAVA levels reduced EAdi and Pdi-time products by 48% (p < 0.05) and 66% (p < 0.05). In contrast, increasing PSV did not reduce the diaphragm electrical activity-time product and increased the transdiaphragmatic pressure-time product (p < 0.05) due to the increased wasted efforts. We conclude that synchrony with the ventilator is an important determinant for diaphragm unloading.Keywords
This publication has 29 references indexed in Scilit:
- Optimal strategies for newborn ventilation—a synthesis of the evidenceEarly Human Development, 2005
- What is the daily practice of mechanical ventilation in pediatric intensive care units? A multicenter studyIntensive Care Medicine, 2004
- Proportional assist ventilation in low birth weight infants with acute respiratory disease: A comparison to assist/control and conventional mechanical ventilationThe Journal of Pediatrics, 1999
- Comparison of Assisted Ventilator Modes on Triggering, Patient Effort, and DyspneaAmerican Journal of Respiratory and Critical Care Medicine, 1997
- Influence of different methods of synchronized mechanical ventilation on ventilation, gas exchange, patient effort, and blood pressure fluctuations in premature neonatesPediatric Pulmonology, 1996
- Patient-Triggered Ventilation Decreases the Work of Breathing in NeonatesAmerican Journal of Respiratory and Critical Care Medicine, 1996
- Respiratory response and inspiratory effort during pressure support ventilation in COPD patientsIntensive Care Medicine, 1995
- Improved oxygenation during synchronized intermittent mandatory ventilation in neonates with respiratory distress syndrome: A randomized, crossover studyThe Journal of Pediatrics, 1995
- The Effectiveness of Pressure Support Ventilation for Mechanical Ventilatory Support in ChildrenAnesthesiology, 1993
- Improved Efficacy of Spontaneous Breathing with Inspiratory Pressure SupportAmerican Review of Respiratory Disease, 1987