Characterization of a fiber‐optic‐coupled radioluminescent detector for application in the mammography energy range
Authors:
- Luis A. BenevidesLBLuis A. Benevides
- Alan L. HustonAHAlan L. Huston
- Brian L. JustusBJBrian L. Justus
- Paul FalkensteinPFPaul Falkenstein
- Libby F. BratemanLBLibby F. Brateman
- David E. HintenlangDHDavid E. Hintenlang
- 25 May 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (6Part1) , 2220-2227
Abstract
Fiber‐optic‐coupled radioluminescent (FOC) dosimeters are members of a new family of dosimeters that are finding increased clinical applications. This study provides the first characterization of a Cu doped quartz FOC dosimeter at diagnostic energies, specifically across the range of x‐ray energies and intensities used in mammographies. We characterize the calibration factors, linearity, angular dependence, and reproducibility of the FOC dosimeters. The sensitive element of each dosimeter was coupled to a photon counting photomultiplier module via long optical fibers. A computer controlled interface permitted real‐time monitoring of the dosimeter output and rapid data acquisition. The axial‐angular responses for all dosimeter models show nearly uniform response without any marked decrease in sensitivity. However, the normal‐to‐axial angular response showed a marked decrease in sensitivity of about and . In most clinical applications, appropriate dosimeter positioning can minimize the contributions of the varying normal‐to‐axial response. The FOC dosimeters having the greatest sensitive length provided the greatest sensitivity, with greatest to lowest sensitivity observed for 4.0, 1.9, 1.6, and length sensitive elements. The average sensitivity of the dosimeters varies linearly with sensitive volume and as a function of tube potential and target/filter combinations, generally exhibiting an increased sensitivity for higher energies. The dosimeter sensitivity as a function of tube potential had an average increase of for dosimeter models and three target‐filter combinations tested (, , and ) over a range of . All dosimeter models exhibited a linear response to exposure for all target‐filter combinations, tube potentials, and tube current‐time product stations evaluated and demonstrated reproducibility within 2%. All of the dosimeters examined in this study provided a response adequate for the accurate measurement of doses in clinical mammography applications.Keywords
Funding Information
- U.S. Department of Energy
This publication has 13 references indexed in Scilit:
- TU‐C‐224C‐09: Radiation Characteristics of a Gated Fiber‐Optic‐Coupled DetectorMedical Physics, 2006
- SU-FF-I-08: Results of An Optical Fiber-Based Dosimetry System for Use in Computed Tomography CharacterizationMedical Physics, 2006
- Characterization of metal oxide semiconductor field effect transistor dosimeters for application in clinical mammographyMedical Physics, 2006
- In vivoabsorbed dose measurements in mammography using a new real-time luminescence techniqueBritish Journal of Radiology, 2005
- A tissue‐equivalent phantom series for mammography dosimetryJournal of Applied Clinical Medical Physics, 2004
- Towardsin vivoTLD dosimetry in mammographyBritish Journal of Radiology, 2004
- Real-time optical-fibre luminescence dosimetry for radiotherapy: physical characteristics and applications in photon beamsPhysics in Medicine & Biology, 2004
- Gated fiber-optic-coupled detector for in vivo real-time radiation dosimetryApplied Optics, 2004
- Estimation of mean-glandular dose from monitoring breast entrance skin air kerma using a high sensitivity metal oxide semiconductor field effect transistor (MOSFET) dosimeter system in mammographyApplied Radiation and Isotopes, 2002
- Remote optical fiber dosimetryNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2001