Comparison of dynamic digital pupillometry and static measurements of pupil size in determining scotopic pupil size before refractive surgery
- 1 June 2005
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 31 (6) , 1171-1176
Abstract
To compare 2 digital infrared pupillometers. Anterior Segment Service, Moorfields Eye Hospital, London, England. Eighty-eight eyes of 44 healthy subjects were recruited into the study. The scotopic pupil size was measured with 2 instruments, a digital pupillometer (Procyon) and a Hartman-Shack wavefront aberrometer (Visx WaveScan). Agreement between the instruments was assessed. There were 21 men and 23 women with a mean age 38.1 years ± 9.2 (SD) (range 23 to 62 years). The mean scotopic pupil diameter was significantly larger with the WaveScan (6.61 ± 0.92 mm) than with the Procyon pupillometer (6.40 ± 0.90 mm), P<.001. There was good agreement in measurement of scotopic pupil size between the two instruments. The larger pupil diameter found with the Visx WaveScan than with the Procyon digital pupillometer probably represents the different illumination level used by each instrument, the effects of accommodation, and target fogging.Keywords
This publication has 14 references indexed in Scilit:
- Pupil size and quality of vision after LASIKOphthalmology, 2003
- Comparison of three different technologies for pupil diameter measurementGraefe's Archive for Clinical and Experimental Ophthalmology, 2003
- Comparison of a digital and a handheld infrared pupillometer for determining scotopic pupil diameterJournal of Cataract & Refractive Surgery, 2003
- Use of a digital infrared pupillometer to assess patient suitability for refractive surgeryJournal of Cataract & Refractive Surgery, 2002
- Comparison of the pupil card and pupillometer in measuring pupil sizeJournal of Cataract & Refractive Surgery, 2002
- Laser in situ keratomileusis with a scanning excimer laser for the correction of low to moderate myopia with and without astigmatismJournal of Cataract & Refractive Surgery, 2001
- Agreement and repeatability of infrared pupillometry and the comparison methodOphthalmology, 1999
- The Effects of Ablation Diameter on the Outcome of Excimer Laser Photorefractive KeratectomyArchives of Ophthalmology (1950), 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983