Quantifying Spurious Free T4 Results Attributable to Thyroxine-Binding Proteins in Serum Dialysates and Ultrafiltrates
Open Access
- 1 May 2007
- journal article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 53 (5) , 985-988
Abstract
Background: Direct equilibrium dialysis and direct ultrafiltration free thyroxine (T4) assays rely on semipermeable membranes to exclude T4-binding serum proteins from dialysates and ultrafiltrates. The presence of these proteins in dialysates or ultrafiltrates will yield spuriously high free T4 values when free T4 is quantified by RIA. Methods: We used a nonanalog free T4 RIA that detects and quantifies dialyzable and ultrafilterable serum free T4 to detect T4-binding serum proteins. Two equilibrium dialysis devices and 3 ultrafiltration devices were used to illustrate this application. Displacements of [125I]T4 from anti-T4 by various concentrations of T4-depleted thyroxine-binding globulin, albumin, and serum total protein were compared to displacements by various concentrations of free T4. Results: Both dialysis devices excluded detectable T4-binding serum proteins from dialysates. Two of 3 ultrafiltration devices excluded detectable T4-binding serum proteins from ultrafiltrates. One did not, and its ultrafiltrate yielded spurious free T4 values that correlated directly with serum protein concentrations. Conclusion: The presence or absence of T4-binding proteins in dialysates and ultrafiltrates and the spurious free T4 values that these proteins cause can be documented using a nonanalog free T4 RIA.Keywords
Funding Information
- Loma Linda University
This publication has 10 references indexed in Scilit:
- Reference methods for the measurement of free thyroid hormones in blood: Evaluation of potential reference methods for free thyroxineClinical Biochemistry, 2004
- Influence of Adsorption and Deproteination on Potential Free Thyroxine Reference MethodsClinical Chemistry, 2002
- Time course of pH regulation in free thyroxin determinations in serumClinical Chemistry, 1991
- New ultrafiltration method for free thyroxin compared with equilibrium dialysis in patients with thyroid dysfunction and nonthyroidal illnessClinical Chemistry, 1990
- Ultrafiltration devices tested for use in a free thyroxine assay validated by comparison with equilibrium dialysisScandinavian Journal of Clinical and Laboratory Investigation, 1990
- Direct determination of free thyroxin in undiluted serum by equilibrium dialysis/radioimmunoassay.Clinical Chemistry, 1988
- Preparation of (T3+T4)-Free Pooled PlasmaAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1986
- Ultrafiltration method for direct radioimmunoassay measurement of free thyroxine and free tri-iodothyronine in serumScandinavian Journal of Clinical and Laboratory Investigation, 1985
- Improved dialysis method for free thyroxin in serum compared with five commercial radioimmunoassays in nonthyroidal illness and subjects with abnormal concentrations of thyroxin-binding globulin.Clinical Chemistry, 1983
- Use of Alkaline Sephadex G-25 for Extraction and Measurement of Serum ThyroxineClinical Chemistry, 1974