Quantification of calcium content in bone by using ToF-SIMS–a first approach
Open Access
- 14 November 2013
- journal article
- Published by American Vacuum Society in Biointerphases
- Vol. 8 (1) , 31
Abstract
The determination of the spatially resolved calcium distribution and concentration in bone is essential for the assessment of bone quality. It enables the diagnosis and elucidation of bone diseases, the course of bone remodelling and the assessment of bone quality at interfaces to implants. With time-of-flight secondary ion mass spectrometry (ToF-SIMS) the calcium distribution in bone cross sections is mapped semi-quantitatively with a lateral resolution of up to 1 μm. As standards for the calibration of the ToF-SIMS data calcium hydroxyapatite collagen scaffolds with different compositions were synthesized. The standards were characterised by loss of ignition, x-ray diffractometry (XRD) and x-ray photoelectron spectroscopy (XPS). The secondary ion count rate for calcium and the calcium content of the standards show a linear dependence. The obtained calibration curve is used for the quantification of the calcium content in the bone of rats. The calcium concentration within an animal model for osteoporosis induction is monitored. Exemplarily the calcium content of the bones was quantified by XPS for validation of the results. Furthermore a calcium mass image is compared with an XPS image to demonstrate the better lateral resolution of ToF-SIMS which advances the locally resolved quantification of the calcium content.Keywords
This publication has 21 references indexed in Scilit:
- Applicability of ToF-SIMS for monitoring compositional changes in bone in a long-term animal modelJournal of The Royal Society Interface, 2013
- Chemical and structural analysis of the bone-implant interface by TOF-SIMS, SEM, FIB and TEM: Experimental study in animalApplied Surface Science, 2012
- Quantification strategies for elemental imaging of biological samples using laser ablation-inductively coupled plasma-mass spectrometryThe Analyst, 2012
- Label free biochemical 2D and 3D imaging using secondary ion mass spectrometryCurrent Opinion in Chemical Biology, 2011
- Human bone material characterization: integrated imaging surface investigation of male fragility fracturesOsteoporosis International, 2011
- Application of Laser Postionization Secondary Neutral Mass Spectrometry/Time-of-Flight Secondary Ion Mass Spectrometry in Nanotoxicology: Visualization of Nanosilver in Human Macrophages and Cellular ResponsesACS Nano, 2011
- Bone mineralization density distribution in health and diseaseBone, 2008
- Clinical Use of Quantitative Computed Tomography and Peripheral Quantitative Computed Tomography in the Management of Osteoporosis in Adults: The 2007 ISCD Official PositionsJournal of Clinical Densitometry, 2008
- The role of DXA bone density scans in the diagnosis and treatment of osteoporosisPostgraduate Medical Journal, 2007
- Major-to-Ultratrace Elements in Bone-Marrow Fluid as Determined by ICP-AES and ICP-MSAnalytical Sciences, 2003