Unexpected sawtooth artifact in beat-to-beat pulse transit time measured from patient monitor data
Authors:
- Yu-Ting LinYLYu-Ting Lin
- Hau-Tieng WuHWHau-Tieng Wu
- Martin G. FraschMartin G. Frasch
Preprint
Abstract
Object: It is increasingly popular to collect as much data as possible in the hospital setting from clinical monitors for research purposes. However, in this setup the data calibration issue is often not discussed and, rather, implicitly assumed, while the clinical monitors might not be designed for the data analysis purpose. We hypothesize that this calibration issue for a secondary analysis may become an important source of artifacts in patient monitor data. We test an off-the-shelf integrated photoplethysmography (PPG) and electrocardiogram (ECG) monitoring device for its ability to yield a reliable pulse transit time (PTT) signal. Approach: This is a retrospective clinical study using two databases: one containing 35 subjects who underwent laparoscopic cholecystectomy, another containing 22 subjects who underwent spontaneous breathing test in the intensive care unit. All data sets include recordings of PPG and ECG using a commonly deployed patient monitor. We calculated the PTT signal offline. Main Results: We report a novel constant oscillatory pattern in the PTT signal and identify this pattern as a sawtooth artifact. We apply an approach based on the de-shape method to visualize, quantify and validate this sawtooth artifact. Significance: The PPG and ECG signals not designed for the PTT evaluation may contain unwanted artifacts. The PTT signal should be calibrated before analysis to avoid erroneous interpretation of its physiological meaning.Keywords
All Related Versions
- Version 1, 2018-08-27, ArXiv
- Version 2, 2019-08-09, ArXiv
- Published version: PLOS ONE, 14 (9), e0221319.
This publication has 16 references indexed in Scilit:
- Multi-sensor fusion in body sensor networks: State-of-the-art and research challengesInformation Fusion, 2017
- Identification of Atrial Fibrillation by Quantitative Analyses of Fingertip PhotoplethysmogramScientific Reports, 2017
- Using Supervised Machine Learning to Classify Real Alerts and Artifact in Online Multisignal Vital Sign Monitoring Data*Critical Care Medicine, 2016
- MIMIC-III, a freely accessible critical care databaseScientific Data, 2016
- Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and PracticeIEEE Transactions on Biomedical Engineering, 2015
- Big data analytics in healthcare: promise and potentialHealth Information Science and Systems, 2014
- Implementation of Artifact Detection in Critical Care: A Methodological ReviewIEEE Reviews in Biomedical Engineering, 2013
- Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based methodEuropean Journal of Applied Physiology, 2011
- Assessment of surgical stress during general anaesthesiaBritish Journal of Anaesthesia, 2007
- Patent Medicine and The ???Black Box???Anesthesia & Analgesia, 2005